Instructions
RIGOL  Chapter 6 Operations and Measurements 
6-24  MSO7000/DS7000 User's Guide 
Auto Measurement 
MSO7000/DS7000 provides auto measurements for 41 waveform parameters, as 
well as the statistics and analysis of the measurement results.   
Quick Measurement after AUTO 
When the oscilloscope is correctly connected and has detected a valid input signal, 
press AUTO to enable the waveform auto setting function and open the auto setting 
function menu. 
  :  Press  this  key,  and  then  the  complete  single  period  of  the  signal  is 
displayed  automatically on  the  screen.  Meanwhile,  the  system  will  make 
measurements  for the  "frequency"  of  the  currently  displayed waveforms  in  a 
complete single period. The measurement results are displayed at the bottom of 
the screen. 
  :  Press  this  key,  and  then  multiple  periods  of  the  signal  are  displayed 
automatically on the screen. Meanwhile, the system will make measurements 
for the "frequency" of the currently displayed waveforms in a multiple periods. 
The measurement results are displayed at the bottom of the screen. 
  :  Press  this  key,  and  then  one  rising  edge  of  the  signal  is  displayed 
automatically on the screen. Meanwhile, the system will make measurements 
for  the  "rise  time"  of  the  currently  displayed  rising  edge.  The  measurement 
results are displayed at the bottom of the screen. By default, it is intended for 
the fast edge signal. 
  :  Press  this  key,  and  then  one  falling  edge  of  the  signal  is  displayed 
automatically on the screen. Meanwhile, the system will make measurements 
for  the  "fall  time"  of  the  currently  displayed  falling  edge.  The  measurement 
results are displayed at the bottom of the screen. By default, it is intended for 
the fast edge signal. 
  : press this key, and the system returns to the last set menu. 
  Press Option to enter the AUTO option setting menu. 
  Press Lock to lock the AUTO function. If the AUTO function is unlocked, 
please perform remote control over the commands. 
  Press Peak to Peak to enable or disable the peak-peak priority setting. 
This function is intended for the shifted signal. If there is a large deviation, 
you can view the signal waveform in priority when you enable the function. 
  Press CH to select "All" or "Open". 
If you select "All", press AUTO and the system will test the 4 analog 
channels (CH1-CH4) in sequence. If no signal is found on the channel, then 
the channel is disabled; if a signal is found on the channel, adjust the 
channel to an optimal scale to display the signal. 
If you select "Open", press AUTO and the system will only test the enabled 