Instructions
RIGOL  Chapter 5 To Trigger the Oscilloscope 
5-54  MSO7000/DS7000 User's Guide 
Trigger Output Connector 
The trigger output connector ([TRIG OUT]) on the rear panel of MSO7000/DS7000 
series can output trigger signals (hardware trigger) determined by the current 
setting. 
Figure 5-23 Trigger Output Connector 
Press Utility  System  Aux OUT to select "TrigOut”. A signal which reflects the 
current capture rate of the oscilloscope can be output from the [TRIG OUT] 
connector each time a trigger is generated by the oscilloscope. If this signal is 
connected to a waveform display device to measure the frequency, the measurement 
result is the same as the current capture rate. 
Note: If you press Utility  System  Aux OUT to select "PassFail", then in the 
pass/fail test, when a pass/fail event is detected, the oscilloscope outputs a pulse 
from the rear-panel [TRIG OUT] connector. 
Trigger Output Connector 