Instructions
RIGOL  Chapter 5 To Trigger the Oscilloscope 
5-32  MSO7000/DS7000 User's Guide 
Nth Edge Trigger 
Triggers on the Nth edge that appears after the specified idle time. For example, in 
the waveform shown in the figure below, the instrument should trigger on the 
second rising edge after the specified idle time (the time between two neighboring 
rising edges), and the idle time should be within the range between P and M (P<Idle 
Time<M). Wherein, M is the time between the first rising edge and its previous rising 
edge; P is the maximum time between the rising edges that participate in counting. 
M
P<Idle Time<M
Edge Type=
Edge Num=2
P
Figure 5-12 Nth Edge Trigger 
Trigger Type: 
Press Type, and then rotate the multifunction knob    to select "Nth Edge". Press 
down the knob to select the trigger type. Then, the current trigger setting 
information is displayed at the upper-right corner of the screen, as shown in the 
figure below. You can also press Type continuously to select the trigger type or 
enable the touch screen to tap the desired trigger type and select it. 
Source Selection: 
Press  Source  to  open  the  signal  source  list  and  select  CH1-CH4  or  D0-D15.  For 
details,  refer  to  descriptions  in  "Trigger  Source".  The  current  trigger  source  is 
displayed at the upper-right corner of the screen. 
Note: Only when we select the channel (that has been input with signals) as the 
trigger source, can we obtain a stable trigger. 
Edge Type: 
Press Slope continuously to select the input signal edge on which the oscilloscope 
triggers. 
  Rising: triggers on the rising edge of the input signal when the voltage level 
meets the specified trigger level. 
  Falling: triggers on the falling edge of the input signal when the voltage level 
meets the specified trigger level. 