Instructions
Chapter 5 To Trigger the Oscilloscope    RIGOL 
MSO7000/DS7000 User's Guide  5-29 
Trigger Mode: 
In the trigger control area (Trigger) on the front panel, press Mode to quickly switch 
the current trigger mode. For details, refer to descriptions in "Trigger Mode". 
Trigger Parameter Setting: 
Set the trigger parameters (trigger holdoff and noise rejection) under  this  trigger 
type.  For  details,  refer  to  descriptions  in  "Trigger  Holdoff"  and  "Noise 
Rejection". 
Trigger LEVEL/Threshold Level: 
Rotate the Trigger    LEVEL knob to adjust the trigger level/threshold level. Refer 
to  "Trigger  LEVEL/Threshold  Level".  The current  trigger  level/threshold  level 
value is displayed at the upper-right corner of the screen. 
Setup/Hold Trigger 
In Setup/Hold trigger, you need to set the clock source and data source. The setup 
time starts when the data signal passes the trigger level and ends at the coming of 
the specified clock edge; the hold time starts at the coming of the specified clock 
edge and ends when the data signal crosses the trigger level again, as shown in the 
figure below. The oscilloscope triggers when the setup time or hold time is smaller 
than the preset time. 
T1
T2
Data Source
Clock Source
T2
T1
is the setup time
is the hold time
The edge type is 
The data type is H
Figure 5-11 Setup/Hold Trigger 
Trigger Type: 
Press Type, and then rotate the multifunction knob    to select "Setup/Hold". 
Press down the knob to select the trigger type. Then, the current trigger setting 
information is displayed at the upper-right corner of the screen, as shown in the 
figure below. You can also press Type continuously to select the trigger type or 
enable the touch screen to tap the desired trigger type and select it. 