Instructions
RIGOL  Chapter 5 To Trigger the Oscilloscope 
5-22  MSO7000/DS7000 User's Guide 
Time Out
T
Time Out< T
Figure 5-8 Timeout Trigger 
Trigger Type: 
Press Type, and then rotate the multifunction knob    to select "Timeout". Press 
down the knob to select the trigger type. Then, the current trigger setting 
information is displayed at the upper-right corner of the screen, as shown in the 
figure below. You can also press Type continuously to select the trigger type or 
enable the touch screen to tap the desired trigger type and select it. 
Source Selection: 
Press  Source  to  open  the  signal  source  list  and  select  CH1-CH4  or  D0-D15.  For 
details,  refer  to  descriptions  in  "Trigger  Source".  The  current  trigger  source  is 
displayed at the upper-right corner of the screen. 
Note: Only when we select the channel (that has been input with signals) as the 
trigger source, can we obtain a stable trigger. 
Edge Type: 
Press Slope and rotate the multifunction knob    to  select the edge type of the 
input signal that starts to pass through the trigger level. You can also press Slope 
continuously to enable the touch screen to select the edge type. 
  Rising: starts timing when the rising edge of the input signal passes through 
the trigger level. 
  Falling:  starts  timing  when  the  falling  edge  of  the  input  signal  passes 
through the trigger level. 
  Either: starts timing when either edge of the input signal passes through the 
trigger level. 
Timeout Value: 
Timeout value represents the maximum time that the signal remains idle before the 
signal passes through  the trigger level. Press Timeout, and then rotate  the 
multifunction knob    or use the pop-up numeric keypad to set the timeout value of 
Timeout trigger. The available range is from 16 ns to 10 s. 