Instructions
Chapter 5 To Trigger the Oscilloscope    RIGOL 
MSO7000/DS7000 User's Guide  5-19 
Trigger Mode: 
In the trigger control area (Trigger) on the front panel, press Mode to quickly switch 
the current trigger mode. For details, refer to descriptions in "Trigger Mode". 
Trigger Parameter Setting: 
Set the trigger parameters (trigger holdoff and noise rejection) under  this  trigger 
type.  For  details,  refer  to  descriptions  in  "Trigger  Holdoff"  and  "Noise 
Rejection". 
Trigger LEVEL/Threshold Level: 
Rotate the Trigger    LEVEL knob to adjust the trigger level/threshold level. Refer 
to  "Trigger  LEVEL/Threshold  Level".  The current  trigger  level/threshold  level 
value is displayed at the upper-right corner of the screen.   
Duration Trigger 
In duration trigger, the instrument identifies a trigger condition by searching for the 
duration of a specified pattern. This pattern is a logical "AND" combination of the 
channels. Each channel can be set to H (high), L (low), or X (don't care). You can 
specify a channel in the pattern to be the rising or falling edge (only one edge can be 
specified). When the duration (△T) of this pattern meets the preset time, the 
oscilloscope initiates a trigger, as shown in the figure below. 
Figure 5-7 Duration Trigger 
Trigger Type: 
Press Type, and then rotate the multifunction knob    to select "Duration". Press 
down the knob to select the trigger type. Then, the current trigger setting 
information is displayed at the upper-right corner of the screen, as shown in the 