Instructions
Chapter 5 To Trigger the Oscilloscope    RIGOL 
MSO7000/DS7000 User's Guide  5-17 
Pattern Trigger 
Identifies a trigger condition by searching for a specified pattern. This pattern is a 
logical "AND" combination of channels. Each channel can be set to H (high), L (low), 
or X (don't care). A rising or falling edge (you can only specify a single edge) can be 
specified for one channel included in the pattern. When an edge is specified, the 
oscilloscope will trigger at the edge specified if the pattern set for the other channels 
are true (namely the actual pattern of the channel is the same as the preset pattern). 
If no edge is specified, the oscilloscope will trigger on the last edge that makes the 
pattern true. If all the channels in the pattern are set to "X", the oscilloscope will not 
trigger. 
Figure 5-5 Pattern Trigger 
Trigger Type: 
Press Type, and then rotate the multifunction knob    to select "Pattern". Press 
down the knob to select the trigger type. Then, the current trigger setting 
information is displayed at the upper-right corner of the screen, as shown in the 
figure below. You can also press Type continuously to select the trigger type or 
enable the touch screen to tap the desired trigger type and select it. 
Source Selection: 
Press  Source  to open  the  signal  source  list  and  select  CH1-CH4  or  D0-D15.  For 
details,  refer  to  descriptions  in  "Trigger  Source".  The  current  trigger  source  is 
displayed at the upper-right corner of the screen. 
Note: Only when we select the channel (that has been input with signals) as the 
trigger source, can we obtain a stable trigger. 
Pattern Setting: 
Press Code, and rotate the multifunction knob    to select the pattern of the 
currently selected channel. Press down the knob to select the pattern. You can also 
press Code continuously to select the pattern or enable the touch screen to select it. 
The available patterns include the following five types: 
  H: sets the pattern of the channel selected to "1", i.g. the voltage level is higher 
than the trigger level/threshold level of the channel. 
  L: sets the pattern of the channel selected to "0", i.g. the voltage level is lower 
than the trigger level/threshold level of the channel. 
  X: sets the pattern of the channel selected to "X", i.g. this channel is not used as 