RIGOL Chapter 2 Command System
2-76 MSO7000/DS7000 Programming Guide
as 0; therefore, the range of the returned value is a decimal number corresponding to a
binary number X0XXXX0X (X is 1 or 0).
Return
Format
The query returns an integer. The integer equals to the binary-weighted sum of all the
bits set in the register.
*IDN?
Syntax
Description
Return
Format
*OPC
Syntax
*OPC
*OPC?
Description
The *OPC command sets bit 0 (Operation Complete, OPC) in the standard event status
register to 1 after the current operation is finished. The OPC? command queries
whether the current operation is finished.
Return
Format
The query returns 1 after the current operation is finished; otherwise, the query returns
0.
*SAV
Syntax
*SAV <value>
Description
Saves the current instrument state to the selected register.
Parameter
Name
Type
Range
Default
<value>
Integer
0 to 49
0
Example
*SAV 1 /*Saves the current instrument state to Register 1.*/