User manual
RIGOL Chapter 2 Command System
2-56 MSO7000/DS7000 Programming Guide
:CURSor:MANual:BYValue?
Syntax
:CURSor:MANual:BYValue?
Description
Queries the Y value at Cursor B in the manual mode of cursor measurement. The unit is
determined by the currently selected vertical unit.
Return
Format
When the channel source is CHANnel1, CHANnel2, CHANnel3, CHANnel4, MATH1,
MATH2, MATH3, or MATH4, the query returns the Y value at Cursor B in scientific
notation.
When the channel source is LA, the query returns a decimal integer that corresponds
to the binary or hex weighted sum of the bits (D15-D0) at Cursor B. For the disabled
channel, the query returns 0 by default.
Example
:CURSor:MANual:BYValue? /*The query returns -2.000000E+0.*/
:CURSor:MANual:XDELta?
Syntax
:CURSor:MANual:XDELta?
Description
Queries the difference (△ X) between the X value at Cursor A and the X value at Cursor B
in the manual mode of cursor measurement. The unit is determined by the currently
selected horizontal unit.
Return
Format
The query returns the current difference in scientific notation.
:CURSor:MANual:IXDELta?
Syntax
:CURSor:MANual:IXDELta?
Description
Queries the reciprocal (1/△ X) of the absolute difference between the X value at Cursor A
and the X value at Cursor B in the manual mode of cursor measurement. The unit is
determined by the currently selected horizontal unit.
Return
Format
The query returns 1/△ X in scientific notation.
:CURSor:MANual:YDELta?
Syntax
:CURSor:MANual:YDELta?
Description
Queries the difference (△ Y) between the Y value at Cursor A and the Y value at Cursor B
in the manual mode of cursor measurement. The unit is determined by the currently
selected vertical unit.
Return
Format
The query returns the current difference value in scientific notation.