User manual

RIGOL Chapter 2 Command System
2-22 DS4000E Programming Guide
Remarks
POSitive: samples the data of the data channel on the rising edge of the clock signal.
NEGative:
samples the data of the data channel on the falling edge of the clock signal.
BOTH: samples the data of the data channel on both the rising and falling edge of the clock signal.
If the clock channel is currently not set (refer to the :BUS<n>:PARallel:CLK command), you do not
need to use the command to set the edge type.
Return Format
The query returns POS, NEG, or BOTH.
:BUS<n>:PARallel:THReshold
Syntax
:BUS<n>:PARallel:THReshold <source>,<thre>
:BUS<n>:PARallel:THReshold? <source>
Description
Sets or queries the threshold level of the specified analog channel of Parallel decoding on the specified bus.
Parameter
Name
Type Range Default
<n> Discrete {1|2} --
<source> Discrete {CHANnel1|CHANnel2|CHANnel3|CHANnel4} CHANnel1
<thre> Real
(-4 x VerticalScale - Offset) to (4 x VerticalScale - Offset)
Wherein, VerticalScale is the vertical scale of the specified
analog channel, refer to the :CHANnel<n>:SCALe command;
Offset is the vertical position of the specified analog channel,
refer to the :CHANnel<n>:OFFSet command.
0
Return Format
The query returns the threshold level in scientific notation.