User manual

RIGOL Chapter 2 Command System
2-160 DS4000E Programming Guide
:MEASure:VMAX:SAVerage
:MEASure:VMAX:SCURrent
:MEASure:VMAX:SDEViation
:MEASure:VMAX:SMAXimum
:MEASure:VMAX:SMINimum
Syntax 1
:MEASure:VMAX:SAVerage [<chan>]
:MEASure:VMAX:SCURrent [<chan>]
:MEASure:VMAX:SDEViation [<chan>]
:MEASure:VMAX:SMAXimum [<chan>]
:MEASure:VMAX:SMINimum [<chan>]
Description 1
Measures the maximum amplitude value of the specified source channel and enables the statistic function
of the measurement results.
Syntax 2
:MEASure:VMAX:SAVerage? [<chan>]
:MEASure:VMAX:SCURrent? [<chan>]
:MEASure:VMAX:SDEViation? [<chan>]
:MEASure:VMAX:SMAXimum? [<chan>]
:MEASure:VMAX:SMINimum? [<chan>]
Description 2
Queries the average value, current value, standard deviation, maximum value, or minimum value of the
measurement results for the maximum amplitude value when the maximum amplitude value measurement
function of the specified source channel and the statistic function of the measurement results are enabled.
The unit is related to the current amplitude unit of the specified source channel.
Parameter
Name
Type
Range
<chan>
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4|MATH}
Remarks
Before making measurements, turn on the channels to be measured.
When the math operation type is FFT (refer to the :CALCulate:MODE command), <chan> cannot be
set to "MATH".
When [<chan>] is omitted, the command, by default, applies to the currently set measurement source
(refer to the :MEASure:SOURce command).
Return Format
The query returns the average value, current value, standard deviation, maximum value, or minimum value
in scientific notation. When the measurement results are invalid, the query returns 9.9e37.