Datasheet

Specications
All the specications (except the parameters marked with "Typical") are guaranteed when the instrument has been working for more than 30
minutes under the specied operating temperature.
Sample
Sample Mode Real-time sample
Max. Real-time Sample Rate 2.0 GSa/s for each channel
Max. Memory Depth 14 Mpts for each channel
Peak Detect 500 ps
Averaging
After all the channels nish N times of sampling at the same time, N can be 2, 4, 8, 16, 32, 64,
128, 256, 512, 1024, 2048, 4096, or 8192.
High Resolution 12 bits when ≥10 μs/div @ 2 GSa/s.
Input
Number of Channels 4 analog channels
Input Coupling DC, AC, or GND
Input Impedance (1 MΩ±1%) || (15 pF±3 pF) or 50 Ω±1.5%
Probe Attenuation Coefcient 0.01X to 1000X, in 1-2-5 step
Maximum Input Voltage (1 MΩ)
CAT I 300 Vrms, CAT II 100 Vrms, transient overvoltage 1000 Vpk
Horizontal
Time Base Scale
DS4024E: 2 ns/div to 1 ks/div
DS4014E: 5 ns/div to 1 ks/div
Deviation between Channels 1 ns (typical), 2 ns (maximum)
Max. Recording Length 14 Mpts for each channel
Time Base Accuracy
[1]
≤ ±4 ppm
Clock Drift ≤ ±2 ppm/year
Delay Range
Pre-trigger (negative delay): Memory Depth/Sample Rate
Post-trigger (positive delay): 1 s to 100 ks
Time Base Mode Y-T, X-Y, Roll, Delayed
Number of X-Ys 2 paths at the same time
Waveform Capture Rate
[2]
60,000 wfms/s
Zero Offset ±0.5 div*minimum time base scale
Vertical
Bandwidth (-3 dB) (50 Ω)
DS4024E: DC to 200 MHz
DS4014E: DC to 100 MHz
Single Bandwidth (50 Ω)
DS4024E: DC to 200 MHz
DS4014E: DC to 100 MHz
Vertical Resolution 8 bits
Vertical Scale
1 MΩ input impedance: 1 mV/div to 5 V/div
50 Ω input impedance: 1 mV/div to 1 V/div
Offset Range
1 MΩ input impedance:
1 mV/div to 225 mV/div: ±2 V
230 mV/div to 5 V/div: ±40 V
50 Ω input impedance:
1 mV/div to 124 mV/div: ±1.2 V
126 mV/div to 1 V/div: ±12 V
Dynamic Range ±5 div
Bandwidth Limit
[1]
DS4024E: 20 MHz/100 MHz
DS4014E: 20 MHz
Low Frequency Response
(AC coupling, -3 dB)
≤5 Hz (on BNC)
RIGOL
4