Datasheet
Specications
All the specications (except the parameters marked with "Typical") are guaranteed when the instrument has been working for more than 30
minutes under the specied operating temperature.
Sample
Sample Mode Real-time sample
Max. Real-time Sample Rate 2.0 GSa/s for each channel
Max. Memory Depth 14 Mpts for each channel
Peak Detect 500 ps
Averaging
After all the channels nish N times of sampling at the same time, N can be 2, 4, 8, 16, 32, 64,
128, 256, 512, 1024, 2048, 4096, or 8192.
High Resolution 12 bits when ≥10 μs/div @ 2 GSa/s.
Input
Number of Channels 4 analog channels
Input Coupling DC, AC, or GND
Input Impedance (1 MΩ±1%) || (15 pF±3 pF) or 50 Ω±1.5%
Probe Attenuation Coefcient 0.01X to 1000X, in 1-2-5 step
Maximum Input Voltage (1 MΩ)
CAT I 300 Vrms, CAT II 100 Vrms, transient overvoltage 1000 Vpk
Horizontal
Time Base Scale
DS4024E: 2 ns/div to 1 ks/div
DS4014E: 5 ns/div to 1 ks/div
Deviation between Channels 1 ns (typical), 2 ns (maximum)
Max. Recording Length 14 Mpts for each channel
Time Base Accuracy
[1]
≤ ±4 ppm
Clock Drift ≤ ±2 ppm/year
Delay Range
Pre-trigger (negative delay): Memory Depth/Sample Rate
Post-trigger (positive delay): 1 s to 100 ks
Time Base Mode Y-T, X-Y, Roll, Delayed
Number of X-Ys 2 paths at the same time
Waveform Capture Rate
[2]
60,000 wfms/s
Zero Offset ±0.5 div*minimum time base scale
Vertical
Bandwidth (-3 dB) (50 Ω)
DS4024E: DC to 200 MHz
DS4014E: DC to 100 MHz
Single Bandwidth (50 Ω)
DS4024E: DC to 200 MHz
DS4014E: DC to 100 MHz
Vertical Resolution 8 bits
Vertical Scale
1 MΩ input impedance: 1 mV/div to 5 V/div
50 Ω input impedance: 1 mV/div to 1 V/div
Offset Range
1 MΩ input impedance:
1 mV/div to 225 mV/div: ±2 V
230 mV/div to 5 V/div: ±40 V
50 Ω input impedance:
1 mV/div to 124 mV/div: ±1.2 V
126 mV/div to 1 V/div: ±12 V
Dynamic Range ±5 div
Bandwidth Limit
[1]
DS4024E: 20 MHz/100 MHz
DS4014E: 20 MHz
Low Frequency Response
(AC coupling, -3 dB)
≤5 Hz (on BNC)
RIGOL