RIGOL Chapter 2 Calibration Procedures
DS4000 Calibration Guide
2-2
Equivalent Calibration
Make sure the oscilloscope has been warmed up for 30 minutes before performing equivalent
calibration. Then, follow the steps below to calibrate the instrument.
1. Disconnect all the channels as well as the external trigger channel. Enter the project mode and
select EqualCal (Figure 4) to start the calibration.
2. The equivalent calibration would last for about 1 minute. Restart the oscilloscope when
“Calibration finished, please restart the oscilloscope!” is displayed.
3. Connect CH1 of the oscilloscope to the [Trig Out/Calibration] connector at the rear panel
using the BNC Cable.
4. Press Utility, enter the second-page menu and press AuxOutput to select “Fast”.
5. Press CH1 and set the input impedance of CH1 to 50 Ω and the vertical scale to 200 mV/div.
Adjust the trigger level to get stable waveform. Adjust the vertical position of CH1 to display the
waveform within the screen range.
6. Observe whether jitter occurs at the signal trigger position at 10 ns and 2 ns horizontal scales
respectively. If yes, calibrate again. The normal calibration waveform should be as shown in Figure
7.
Figure 7