User manual
Appendix Test Record Form RIGOL
MSO4000/DS4000 Performance Verification Guide
1
Appendix Test Record Form
RIGOL MSO4000/DS4000 Series Digital Oscilloscope
Performance Verification Test Record Form
Model:
Tested by:
Test Date:
Impedance Test
CH1-CH4 (1 MΩ Input Impedance)
Channel
Vertical Scale
Test Result
Limit
Pass/Fail
CH1
100 mV/div
0.99 MΩ to 1.01 MΩ
500 mV/div
CH2
100 mV/div
500 mV/div
CH3
100 mV/div
500 mV/div
CH4
100 mV/div
500 mV/div
CH1-CH4 (50 Ω Input Impedance)
Channel
Vertical Scale
Test Result
Limit
Pass/Fail
CH1
100 mV/div
49.25 Ω to 50.75 Ω
500 mV/div
CH2
100 mV/div
500 mV/div
CH3
100 mV/div
500 mV/div
CH4
100 mV/div
500 mV/div
External Trigger Channel
Channel
Input Impedance
Test Result
Limit
Pass/Fail
EXT TRIG
1 MΩ 0.99 MΩ to 1.01 MΩ
50 Ω 49.25 Ω to 50.75 Ω