User manual

RIGOL Chapter 2 Performance Verification Test
MSO4000/DS4000 Performance Verification Guide
2-14
point, the amplitude loss should be lower than -3 dB.
15) Keep the other settings of the oscilloscope in step 3) unchanged and set the vertical scale to
200 mV/div.
16) Output a Sine with 1 MHz frequency and 1.2 Vpp amplitude via Fluke 9500B.
17) Repeat step 6).
18) Output a Sine with 200 MHz frequency and 1.2 Vpp amplitude via Fluke 9500B.
19) Repeat steps 8) to 10).
20) Output a Sine with 400 MHz frequency and 1.2 Vpp amplitude via Fluke 9500B.
21) Repeat steps 12) to 14).
22) Keep the other settings of the oscilloscope in step 3) unchanged and set the vertical scale to
500 mV/div.
23) Output a Sine with 1 MHz frequency and 3 Vpp amplitude via Fluke 9500B.
24) Repeat step 6).
25) Output a Sine with 200 MHz frequency and 3 Vpp amplitude via Fluke 9500B.
26) Repeat steps 8) to 10).
27) Output a Sine with 400 MHz frequency and 3 Vpp amplitude via Fluke 9500B.
28) Repeat steps 12) to 14).
29) Turn off CH1. Test CH2, CH3 and CH4 respectively using the method above.