Chapter 2 Performance Verification Test
This chapter introduces the performance verification test methods and procedures of
MSO4000/DS4000 series digital oscilloscope by taking MSO4054 as an example. Fluke 9500B is used
in this manual for the tests. You can also use other devices that fulfill the Specificationin Table 2-1.
Table 2-1 Test Devices Required
Test Plan Device Specification Recommended Model
Plan 1
Oscilloscope
Calibrator
DC output voltage range:
1 : 1 mV to 200 V
50 MΩ: 1 mV to 200 V
Fast edge signal rise time: 150 ps
Fluke 9500B
Plan 2
Digital Multimeter
The resistance measurement
accuracy is higher than ±0.1% of
reading
RIGOL DM3058/3068
Test Cable
BNC (M)-Dual banana plug cable
--
Function/Arbitrary
Waveform
Generator
Frequency accuracy: ±1 ppm
RIGOL DG4162
Dual-BNC Cable
BNC (M)-BNC (M) cable
--
Note:
1. Make sure that the oscilloscope passes the self-test and self-calibration is performed before
executing the performance verification tests.
2. Make sure that the oscilloscope has been warmed up for at least 30 minutes before executing any
of the following tests.
3. Please reset the instrument to the factory setting before or after executing any of the following
tests.