User manual
Chapter 2 Command System RIGOL
DS4000E Programming Guide 2-39
:CALCulate:ADD
Command List:
:CALCulate:ADD:INVert
:CALCulate:ADD:SA
:CALCulate:ADD:SB
:CALCulate:ADD:VOFFset
:CALCulate:ADD:VSCale
:CALCulate:ADD:INVert
Syntax
:CALCulate:ADD:INVert <bool>
:CALCulate:ADD:INVert?
Description
Enables or disables the invert display of the waveform of addition operation; or queries the on/off status of
the invert display of the waveform of addition operation.
Parameter
Name Type Range Default
<bool> Bool {{1|ON}|{0|OFF}} 0|OFF
Remarks
1|ON: turns the waveform invert on. The waveform voltage values are inverted and then displayed.
0|OFF: turns the waveform invert off. The waveform is displayed normally.
Return Format
The query returns 1 or 0.
:CALCulate:ADD:SA
:CALCulate:ADD:SB
Syntax
:CALCulate:ADD:SA <source>
:CALCulate:ADD:SA?
:CALCulate:ADD:SB <source>
:CALCulate:ADD:SB?
Description
Sets or queries the channel source of Source A or Source B in "add" operation.
Parameter
Name
Type Range Default
<source> Discrete {CHANnel1|CHANnel2|CHANnel3|CHANnel4} CHANnel1
Return Format