User Manual
RIGOL 1 Quick Start
DS4000 User’s Guide
1-18
AUTO
Press this key to enable the waveform auto setting
function. The oscilloscope will automatically adjust the
vertical scale, horizontal time base and trigger mode
according to the input signal to realize optimum waveform
display. Besides, quick parameter measurement function
is also provided (for more details, refer to “Quick
Measurement after AUTO”).
Note that auto setting requires that the frequency of the
signal under test should be no lower than 50 Hz, the duty
cycle be greater than 1% and the amplitude be at least 20
mVpp. If the parameters exceed these limits, “Auto failed”
would be displayed after pressing this key and the quick
parameter measurement menu might not be displayed.
RUN/STOP
Press this key to set the state of the oscilloscope to
“RUN” or “STOP”. In “RUN” state, the key is illuminated in
yellow and red in “STOP” state.
SINGLE
Press this key to set the trigger mode to “Single”. In
single trigger mode, the oscilloscope generates a trigger
when the trigger conditions are met and then stops.