Analysis Based on Pass/Fail Mask
Press Analyze and select “Pass/Fail”. Then, set the templet used in analysis based
on Pass/Fail mask through the method below.
1. X Mask
Press this softkey and use to adjust the horizontal threshold and the range
is from 0.02 div to 4.00 div. During the adjustment, two curves would be
displayed to show the outline of the current mask as shown in the figure below.
2. Y Mask
Press this softkey and use to adjust the vertical threshold and the range is
from 0.03 div to 4.00 div. During the adjustment, two curves would be displayed
to show the outline of the current mask.
3. Create Mask
Mask is the templet used in waveform analysis. Press Create Mask to
immediately apply the mask (X Mask and Y Mask) currently created.
4. Save
Press Save to enter the file store interface. Store the test mask file to internal or
external memory by referring to the introduction in Store and Recall.
5. Load
Press Load to enter the file recall interface. Load the test mask into the internal
memory of the instrument by referring to the introduction in Store and
Recall.