RIGOL 9 Pass/Fail Test
DS4000 Users Guide
9-4
Test and Ouput
Before the test, you can use the following method to set the output mode of the test
results.
Press Msg Display to select ON or OFF. When ON is selected, the test results
will be displayed at the upper right corner of the screen.
Press Stop On Outp to select ON or OFF.
ON: when failed waveforms are detected, the oscilloscope will stop the test and
enter the STOP state. At this point, the results of the test remain the same on
the screen (if display is turned on) and only one pulse is output from the [Trig
Out/Calibration] (if enabled) connector at the rear panel.
OFF: the oscilloscope will continue with the test even though failed waveforms
are detected. The test results on the screen will update continuously and a pulse
will be output from the [Trig Out/Calibration] connector at the rear panel
each time a failed waveform is detected.
Press Output to select “Fail” or “Fail+ ”.
Fail: when failed waveforms are detected, there are display and output but the
beeper does not sound.
Fail+ : when failed waveforms are detected, there are display and output and
the beeper sounds (not related to the on/off state of the sound).
Press AuxOutput to quickly turn ON or OFF the output of test results from the
[Trig Out/Calibration] connector at the rear panel. You can also press Utility
AuxOutput and select PassFail to set this output.