Address: trigger on the clock (SCL) edge corresponding to the byte of data (SDA)
behind the preset address (Write, Read or R/W direction). After this trigger
condition is selected:
--press AddrBits to select “7 bit” or “10 bit”;
--press Address to set the address value according to the setting in AddrBits
and the ranges are from 0 to 127 and from 0 to 1023 respectively;
--press Direction to select “Read”, “Write” or “R/W”.
Data: the trigger searches for the control byte value on the data line (SDA)
following which there is a reading bit and an acknowledgement bit and then
searches for the specified data value and qualifier. When this event occurs, the
oscilloscope will trigger on the clock edge of the acknowledgement bit behind
the data byte. After this trigger condition is selected:
--press Byte Length to set the length of the data and the range is from 1 to 5;
--press CurrentBit to select the desired data bit and the range is from 0 to
(Byte Length× 8-1);
--press Data to set the data pattern of the current data bit to X, H or L.
--press AllBits to set the data pattern of all the data bits to the data pattern
specified in Data.
A&D: trigger when the Address” and “Data” conditions are met at the same
time. After this trigger condition is selected:
--press AddrBits to select “7 bit” or “10 bit”;
--press Address to set the address value according to the setting in AddrBits
and the ranges are from 0 to 127 and from 0 to 1023 respectively;
--press Byte Length to set the length of the data and the range is from 1 to 5;
--press CurrentBit to select the desired data bit and the range is from 0 to
(Byte Length× 8-1);
--press Data to set the data pattern of the current data bit to X, H or L.
--press AllBits to set the data pattern of all the data bits to the data pattern
specified in Data.
--press Direction to select “Read”, “Write” or “R/W”.
Trigger Mode:
Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto,
normal or single. The corresponding status light of the current trigger mode turns on.