RIGOL User’s Guide DS4000 Series Digital Oscilloscope Nov. 2011 RIGOL Technologies, Inc.
RIGOL Guaranty and Declaration Copyright © 2011 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information RIGOL is a registered trademark of RIGOL Technologies, Inc. Publication Number UGA10103-1110 Notices RIGOL products are protected by patent law in and outside of P.R.C. RIGOL reserves the right to modify or change parts of or all the specifications and pricing policies at company’s sole decision. Information in this publication replaces all previously corresponding material.
RIGOL Safety Requirement General Safety Summary Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injuries or damages to the instrument and any product connected to it. To prevent potential hazards, please use the instrument as specified in this manual. Use Proper Power Cord. Only power cord designed for the instrument and authorized by local country could be used. Ground The Instrument.
RIGOL Avoid Circuit or Wire Exposure. Do not touch exposed junctions and components when the unit is powered. Do Not Operate With Suspected Failures. If you suspect damage occurs to the instrument, have it inspected by qualified service personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel. Keep Well Ventilation. Inadequate ventilation may cause increasing of temperature or damages to the device.
RIGOL Safety Terms and Symbols Terms in this Manual. The following terms may appear in this manual: WARNING Warning statements indicate the conditions or practices that could result in injury or loss of life. CAUTION Caution statements indicate the conditions or practices that could result in damage to this product or other property. Terms on the Product. These terms may appear on the product: DANGER indicates an injury or hazard may immediately happen.
RIGOL Measurement Category Measurement Category DS4000 series digital oscilloscopes can make measurements in Measurement Category I. WARNING This oscilloscope can only be used for measurements within its specified measurement categories. Measurement Category Definitions Measurement category I is for measurements performed on circuits not directly connected to MAINS. Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS derived circuits.
RIGOL Ventilation Requirement This oscilloscope uses fan to force cooling. Please make sure that the air intake and exhaust areas are free from obstructions and have free air. When using the oscilloscope in a bench-top or rack setting, provide at least 10 cm clearance beside, above and behind the instrument for adequate ventilation. WARNING Inadequate ventilation may cause temperature increase which would damage the instrument.
RIGOL Working Environment Temperature Operating: 0℃ to +50℃ Non-operating: -20℃ to +70℃ Humidity Under +35℃: ≤90% relative humidity +35℃ to +40℃: ≤60% relative humidity WARNING To avoid short circuit inside the instrument or electric shock, please do not operate in humid environment. Altitude Operating: less than 3 km Non-operating: less than 15 km Installation (overvoltage) Category This product is powered by mains conforming to installation (overvoltage) category II.
RIGOL Pollution Degree Degree 2 Pollution Degree Definitions Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no influence. For example: a clean room or air-conditioned office environment. Pollution degree 2: Normally only dry, non-conductive pollution occurs. Occasionally a temporary conductivity caused by condensation may occur. For example: general indoor environment.
RIGOL General Care and Cleaning General Care: Do not store or leave the instrument at places where the instrument will be exposed to direct sunlight for long periods of time. Cleaning: Clean the instrument regularly according to its operating conditions. To clean the exterior surface: 1. Disconnect the instrument from all power sources. 2. Clean the loose dust on the outside of the instrument with a lint- free cloth (with mild detergent or water). When cleaning the LCD, take care to avoid scarifying it.
RIGOL Environmental Considerations The following symbol indicates that this product complies with the applicable European Union requirements according to Directives 2002/96/EC on waste electrical and electronic equipment (WEEE). Product End-of-Life Handling The equipment may contain substances that could be harmful to the environment or human health.
RIGOL DS4000 Series Overview Being a multifunctional and high performance digital oscilloscope, DS4000 provides superb specifications and various functions, which in combination with its easy-to-use design can help users to fulfill their tasks (such as measurement and remote control) more quickly. Main features: 500 MHz, 350 MHz, 200 MHz and 100 MHz bandwidth. Dual-channel or 4-channel model. 4 GSa/s maximum real-time sample rate and 110,000 wfms/s (dots display) maximum waveform refresh rate.
RIGOL Document Overview 1 Quick Start Provide information about preparations before using the instrument and a brief introduction of the instrument. 2 To Set the Vertical System Introduce the functions of the vertical system of the oscilloscope. 3 To Set the Horizontal System Introduce the functions of the horizontal system of the oscilloscope. 4 To Set the Sample System Introduce the functions of the sample system of the oscilloscope.
RIGOL oscilloscope. 13 System Function Setting Introduce how to set the remote interface and system-related functions. 14 Remote Control Introduce how to control the oscilloscope remotely. 15 Troubleshooting Introduce how to deal with common failures of the oscilloscope. 16 Specifications List the specifications and general specifications of the oscilloscope. 17 Appendix Provide common information such as options and accessories.
RIGOL Content Conventions in this Manual: This manual takes DS4054 for example and the descriptions here have contained all the functions and performances of other models.
RIGOL Contents Guaranty and Declaration .........................................................................I Safety Requirement ................................................................................ II General Safety Summary ........................................................................... II Safety Terms and Symbols ....................................................................... IV Measurement Category ......................................................................
RIGOL Multi-function Knob........................................................................ 1-19 Navigation Knob ............................................................................ 1-19 Default ......................................................................................... 1-20 Print ............................................................................................. 1-20 Function Keys................................................................................
RIGOL Sample Rate ......................................................................................... 4-6 Memory Depth ...................................................................................... 4-8 Antialiasing ........................................................................................... 4-9 5 To Trigger the Oscilloscope ................................................................ 5-1 Trigger Source ........................................................................
RIGOL To Clear the Measurement .............................................................. 6-23 All Measurement............................................................................ 6-24 Statistic Function ........................................................................... 6-25 Measurement History ..................................................................... 6-26 Cursor Measurement ............................................................................ 6-27 Manual Mode .......
RIGOL Analysis Based on Pass/Fail Mask................................................... 10-13 11 Display Control............................................................................... 11-1 To Select the Display Type ..................................................................... 11-2 To Set the Persistence Time .................................................................. 11-3 To Set the Waveform Brightness ............................................................
RIGOL External Trigger Impedance .......................................................... 13-13 Aux Output ................................................................................. 13-14 Reference Clock........................................................................... 13-15 Option Management .................................................................... 13-16 14 Remote Control .............................................................................
1 Quick Start RIGOL 1 Quick Start This chapter introduces the preparations when using the oscilloscope for the first time, the front panel, rear panel and user interface of the oscilloscope as well as the using method of the built-in help system.
RIGOL 1 Quick Start General Inspection 1. Inspect the shipping container for damage. Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has passed both electrical and mechanical tests. The consigner or carrier shall be liable for the damage to instrument resulting from shipment. RIGOL would not be responsible for free maintenance/rework or replacement of the unit. 2. Inspect the instrument.
1 Quick Start RIGOL Appearance and Dimensions Figure 1-1 Front View 218.6 207 440 Unit: mm 130.7 116.
1 Quick Start RIGOL To Prepare the Oscilloscope for Use To Remove the Cover Before using the oscilloscope, remove the front panel cover by grasping the transverse grab on each side and pull them in the arrow directions as shown in the figure below.
1 Quick Start RIGOL To Adjust the Supporting Legs Adjust the supporting legs properly to use them as stands to tilt the oscilloscope upwards for stable placement of the oscilloscope as well as better operation and observation. Unfold or fold the supporting legs in the arrow directions as shown in the figures below.
1 Quick Start RIGOL To Connect to AC Power Supply This oscilloscope accepts two kinds of AC power supplies: 100-127V, 45-440Hz and 100-240V, 45-65Hz. Please use the power cord supplied with the accessories to connect the oscilloscope to the power source as shown in the figure below. Turn on the power switch under the power plug; at this point, the oscilloscope is energized and the power key at the lower-left corner of the front panel is in breathing state.
1 Quick Start RIGOL Power-on Inspection When the oscilloscope is energized, press the power key at the lower-left corner of the front panel to start the oscilloscope. During the start-up process, the oscilloscope performs a series of self-tests and you can hear the sound of relay switching. After the self-test, the start-up image is displayed. Press Utility System SelfTestInfo to view the self-test results.
1 Quick Start RIGOL To Connect the Probe RIGOL provides passive and active probes for the DS4000 series oscilloscopes. For detailed technical information of the probes, please refer to the corresponding Probe User’s Guide. The following are the probes recommended for this oscilloscope. Model Description RP3500 500 MHz, passive probe, standard, auto detection RP7150 1.5 GHz, active probe, optional, auto detection Connect the Probe: 1.
1 Quick Start RIGOL Function Inspection 1. Press Default to restore the oscilloscope to its default configuration. 2. Connect the ground alligator clip of the probe to the “Ground Terminal” under the probe compensation signal output terminal. 3. Use the probe to connect the input terminal of channel 1 (CH1) of the oscilloscope and the “Compensation Signal Output Terminal” of the probe. Compensation Signal Output Terminal Ground Terminal Figure 1-7 To Use the Compensation Signal 4. Press AUTO. 5.
RIGOL 1 Quick Start Tip The signal output from the probe compensation connector can only be used for probe compensation adjustment and can not be used for calibration.
1 Quick Start RIGOL Probe Compensation When the probes are used for the first time, you should compensate the probes to match the input channels of the oscilloscope. Non-compensated or poorly compensated probes may cause measurement inaccuracy and error. The probe compensation procedures are as follows: 1. Perform steps 1, 2, 3 and 4 of “Function Inspection” in the previous section. 2. Check the waveforms displayed and compare them with the following.
1 Quick Start RIGOL Front Panel Overview (1) (2) (3) (4) (5) (6) (7) (8) (9) (10) (11) (12) (13) (14) (15) (16) (17) (18) (19) (20) Figure 1-10 Front Panel Overview Table 1-1 Front Panel Description No. Description No.
1 Quick Start RIGOL Rear Panel Overview Trig Out/Calibration Video Output Reference Clock 探头校准 时钟 输出 LAN USB HOST AC Input/Switch USB DEVICE Lock Hole DEVICE HOST Figure 1-11 Rear Panel Overview 1. Trig Out/Calibration Various kinds of signals can be output from this connector (press Utility AuxOutput to select the desired output type). 1) TrigOut: the oscilloscope outputs a signal which can reflect the current capture rate of the oscilloscope each time a trigger is generated.
RIGOL 1 Quick Start synchronize two or more oscilloscope clocks. 3. Video Output Through this interface, the oscilloscope can be connected to external monitors to get clearer waveform display. Note that the display of the oscilloscope is still valid. 4. LAN Through this interface, the oscilloscope can be connected to the local area network for remote control. As the oscilloscope conforms to the LXI-C instrument standards, a test system can be built quickly and accessed through webpage. 5.
1 Quick Start RIGOL Front Panel Function Overview VERTICAL DS4000 provides independent vertical control systems for the four analog input channels. CH1, CH2, CH3, CH4: analog input channels. The four channels are marked by different colors which are also used to mark both the corresponding waveforms on the screen and the channel input connectors. Press any key to open the corresponding channel menu and press again to turn off the channel. AC: illuminated when AC channel coupling is enabled.
RIGOL 1 Quick Start decrease the scale with a step of 5-2-1 and turn counterclockwise to increase with a step of 1-2-5. Press down the knob to switch the vertical scale adjustment mode between “Coarse” and “Fine”. Decode1 and Decode2: decoding function keys. Press the corresponding key to open the decoding function menu. DS4000 supports parallel decoding and protocol decoding (for more details, refer to “Protocol Decoding”).
1 Quick Start RIGOL TRIGGER MODE: press this key to switch the trigger mode to Auto, Normal or Single and the state backlight of the current trigger mode would be illuminated. TRIGGER LEVEL: modify the trigger level. Turn clockwise to increase the level and turn counterclockwise to reduce. During the modification, the trigger level line moves up and down and the value in the trigger level message box (such as ) at the lower-left corner of the screen would change accordingly.
RIGOL 1 Quick Start AUTO Press this key to enable the waveform auto setting function. The oscilloscope will automatically adjust the vertical scale, horizontal time base and trigger mode according to the input signal to realize optimum waveform display. Besides, quick parameter measurement function is also provided (for more details, refer to “Quick Measurement after AUTO”).
1 Quick Start RIGOL Multi-function Knob This knob can be used to adjust the value of the parameters, adjust the waveform brightness and select the desired menu (the backlight turns on during operation). During parameter input, turn the knob clockwise to increase the parameter and counterclockwise to decrease. When the menu is hidden, rotate the knob to adjust the waveform brightness. The adjustable range is from 0% to 100%.
RIGOL 1 Quick Start Default Press this key to restore the oscilloscope to its default configuration. Print Press this key to execute print function or save the screen in the USB storage device. If the oscilloscope is currently connected to a printer and the printer is in idle state, press this key to execute print function. If no printer but a USB storage device is currently connected, press this key to save the screen to the USB storage device in “.bmp” format.
1 Quick Start RIGOL Display: press this key to enter display setting menu to set the display type, persistence time, brightness, grid type, grid brightness and menu display time of the waveform. Utility: press this key to enter system function setting menu to set the system-related functions or parameters, such as I/O setting, sound and language. The oscilloscope also supports some advanced functions such as pass/fail test and print setting.
1 Quick Start RIGOL User Interface DS4000 oscilloscope provides 9 inches, WVGA (800*480) 160,000 color TFT LCD. What is worth mentioning is that the 14-grid ultra-wide screen enables you to view “longer” waveform.
1 Quick Start RIGOL 1. Measurement Menu Provide 12 horizontal (HORIZONTAL) and 10 vertical (VERTICAL) measurement parameters. Press the softkey at the left of the screen to activate the corresponding measurement item. 2. Channel Label/Waveform Different channels are marked by different colors and the color of the waveform complies with the color of the channel. 3. Status Available states include RUN, STOP, T’D (triggered), WAIT and AUTO. 4.
1 Quick Start RIGOL the knob to automatically set the parameter to 0. 9. Trigger Type Display the currently selected trigger type and trigger condition setting. Different label is displayed when different trigger type is selected. For example, represents triggering on the rising edge in “Edge” trigger. 10. Trigger Source Display the trigger source (CH1 to CH4, EXT, EXT/5 or AC Line) currently selected.
1 Quick Start (such as RIGOL when the input impedance is 50 Ω). 14. CH3 Vertical Scale Display the voltage value per grid of CH3 waveform vertically. Use VIRTICAL SCALE to modify this parameter. The following labels will be provided according to the current channel setting: channel coupling (such as in AC coupling), bandwidth limit (such as when bandwidth limit is enabled) and input impedance (such as when the input impedance is 50 Ω). 15.
1 Quick Start RIGOL 18. Operation Menu Press any softkey to activate the corresponding menu. The following symbols might be displayed in the menu: Denote that at the front panel can be used to select parameter items. The backlight of turns on when parameter selection is valid. Denote that can be used to modify parameter value. The backlight of turns on when parameter input is valid. Denote that you can use the navigation knob to quickly adjust/locate parameters.
1 Quick Start RIGOL To Use the Security Lock If needed, you can use the security lock (please buy it yourself) to lock the oscilloscope to a fixed location. The method is as follows, align the lock with the lock hole and plug it into the lock hole vertically, turn the key clockwise to lock the oscilloscope and then pull the key out. Security Lock Hole Figure 1-13 To Use the Security Lock Note: do not insert other articles into the security lock hole to avoid damaging the instrument.
1 Quick Start RIGOL To Use the Built-in Help System The help system of this oscilloscope provides instructions for all the function keys (including menu keys) at the front panel. Press Help to open the help interface and press again to close the interface. The help interface mainly consists of two parts. The left is “Help Options” and you can use “Button” or “Index” mode for selection. The right is “Help Display Area”. Help Options Help Display Area Figure 1-14 Help Information Button: Default mode.
2 To Set the Vertical System RIGOL 2 To Set the Vertical System The contents of this chapter: To Enable the Channel Channel Coupling Bandwidth Limit Probe Input Impedance Waveform Invert Vertical Scale Vertical Expansion Amplitude Unit Channel Label Delay Calibration DS4000 User’s Guide 2-1
2 To Set the Vertical System RIGOL To Enable the Channel DS4000 provides four analog input channels (CH1 to CH4) and provides independent vertical control system for each channel. As the vertical system setting methods of the four channels are completely the same, this chapter takes CH1 as an example to introduce the setting method of the vertical system.
2 To Set the Vertical System RIGOL Channel Coupling Set the coupling mode to filter out the undesired signals. For example, the signal under test is a square waveform with DC offset. When the coupling mode is “DC”: the DC and AC components of the signal under test can both pass the channel. When the coupling mode is “AC”: the DC components of the signal under test are blocked. When the coupling mode is “GND”: the DC and AC components of the signal under test are both blocked.
RIGOL 2 To Set the Vertical System Bandwidth Limit Set the bandwidth limit to reduce display noise. For example, the signal under test is a pulse with high frequency oscillation. When bandwidth limit is disabled, the high frequency components of the signal under test can pass the channel. Enable bandwidth limit and limit the bandwidth to 20 MHz, 100 MHz or 200 MHz, the high frequency components that exceed 20 MHz, 100 MHz or 200 MHz are blocked.
2 To Set the Vertical System RIGOL Probe This oscilloscope supports normal passive probe and active differential probe and can automatically identify the type of the probe currently connected and the probe ratio. Press CH1 Probe to open the probe operation menu. 1. ProbeType Read the type of the probe currently connected as “Nor-Probe” or “DiffProbe”. Note that when a 50 Ω “DiffProbe” is used, the Input Impedance of the channel is set to “50 Ω” automatically. Normal Probe: such as RIGOL RP3500.
RIGOL 2 To Set the Vertical System 3. Front-End RP7150 active probe provides “Single-end” and “Difference” probe heads. Press this softkey to select the desired probe head. 4. Probe-Cal Connect the differential probe to the channel input terminal (such as CH1) of the oscilloscope correctly, then connect the fast edge signal output from the [Trig Out/Calibration] connector at the rear panel of the instrument to the probe and CH2 respectively.
2 To Set the Vertical System RIGOL Input Impedance To reduce the circuit load caused by the interaction between the oscilloscope and the circuit under test, the oscilloscope provides two input impedance modes: 1 MΩ (default) and 50 Ω. 1 MΩ: at this point, the input impedance of the oscilloscope is very high and the current flows into the oscilloscope from the circuit under test can be ignored. 50 Ω: match the oscilloscope with devices with 50 Ω output impedance.
RIGOL 2 To Set the Vertical System Waveform Invert When waveform invert is enabled, the waveform display rotates 180 degree relative to the ground potential. When waveform invert is disabled, the waveform display is normal. Press CH1 Invert to enable or disable waveform invert. Vertical Scale The vertical scale can be adjusted in “Coarse” or “Fine” mode. Press CH1 Volts/Div to select the desired mode.
2 To Set the Vertical System RIGOL Vertical Expansion When using VERTICAL SCALE to change the vertical scale of the analog channel, you can choose to expand or compress the signal vertically around the center of the screen or the ground point of the signal. Press Utility System VerticalExp to select “Center” or “Ground” and the default is “Ground”. Center: when the vertical scale is modified, the waveform will expand or compress around the center of the screen.
2 To Set the Vertical System RIGOL Channel Label You can modify the labels used to mark the analog channels (CH1 to CH4) at the left side of the screen. The label is the number of the channel (such as ) by default and the length of the label can not exceed 4 characters. Press CH1 Label to enter the label modification interface as shown in the figure below. Name Input Area For example, change “ Keyboard ” to “ Upper/Lower Case Switch ”. Press Keyboard to select the “Keyboard” area.
2 To Set the Vertical System RIGOL Delay Calibration When using an oscilloscope for actual measurement, the transmission delay of the probe cable may bring greater error (zero offset). DS4000 allows users to set a delay time to calibrate the zero offset of the corresponding channel. Zero offset is defined as the offset of the crossing point of the waveform and trigger level line relative to the trigger position, as shown below.
3 To Set the Horizontal System RIGOL 3 To Set the Horizontal System The contents of this chapter: Delayed Sweep Time Base Mode Horizontal Scale Horizontal Reference DS4000 User’s Guide 3-1
3 To Set the Horizontal System RIGOL Delayed Sweep Delayed sweep can be used to enlarge a length of waveform horizontally to view the waveform details. MENU Press MENU in the horizontal control area (HORIZONTAL) and press Delayed to enable or disable delayed sweep. Note that to enable delayed sweep, the current time base mode must be “Y-T” and the “Pass/Fail test” must be disabled.
3 To Set the Horizontal System RIGOL In delayed sweep mode, the screen is divided into two display areas as shown in the figure below. The waveform before enlargement Main time base Delayed Sweep Time Base The waveform after enlargement The waveform before enlargement: The waveform in the area that has not been covered by the subtransparent blue in the upper part of the screen is the waveform before enlargement.
RIGOL 3 To Set the Horizontal System Time Base Mode Press MENU in the horizontal control area at the front panel and then press Time Base to select the time base mode of the oscilloscope and the default is Y-T. Y-T Mode This mode is the main time base mode and is applicable to CH1 to CH4. In this mode, the Y axis represents voltage and the X axis represents time. Note that only when this mode is enabled can delayed sweep be turned on.
3 To Set the Horizontal System RIGOL X-Y Mode In this mode, the oscilloscope automatically turns on all the four channels (CH1 to CH4) and the screen is divided into two coordinate areas, wherein, X1 and Y1 track the voltages of CH1 and CH2 and X2 and Y2 track the voltages of CH3 and CH4. The phase deviation between two signals with the same frequency can be easily measured via Lissajous method. The figure below shows the measurement schematic diagram (also called “Ellipse Method”) of the phase deviation.
RIGOL 3 To Set the Horizontal System X-Y function can be used to measure the phase deviation occurred when the signal under test passes through a circuit network. Connect the oscilloscope to the circuit to monitor the input and output signals of the circuit. Application example: measure the phase deviation of the input signals of two channels. Method 1: Use Lissajous method 1. 2. 3. 4.
3 To Set the Horizontal System RIGOL When X-Y mode is enabled, delayed sweep will be disabled automatically. The following functions are not available in X-Y mode: Auto measure, cursor measure, math operation, reference waveform, delayed sweep, vector display, HORIZONTAL POSITION, trigger control, memory depth, acquisition mode, Pass/Fail test and waveform record.
RIGOL 3 To Set the Horizontal System Roll Mode In this mode, the waveform scrolls from the right to the left to update the display and the waveform horizontal position and trigger control are not available. The range of horizontal scale adjustment is from 200.0 ms to 50.00 s. Note: when Roll mode is enabled, “Delayed Sweep”, “Protocol Decoding”, “Pass/Fail Test”, “Measurement Range”, “Waveform Record”, “To Set the Persistence Time” and “To Trigger the Oscilloscope” are not available.
3 To Set the Horizontal System RIGOL Horizontal Scale Being similar to “Vertical Scale”, the horizontal scale can be adjusted in “Coarse” or “Fine” mode. Press MENU ScaleAdjust in the horizontal control area (HORIZONTAL) at the front panel to select the desired mode. Turn HORIZONTAL SCALE to adjust the horizontal scale. Turn clockwise to reduce the horizontal scale and turn counterclockwise to increase.
RIGOL 3 To Set the Horizontal System Horizontal Reference Horizontal reference is the reference position according to which the screen waveform expands and compresses horizontally when adjusting HORIZONTAL SCALE. In Y-T mode (this function is not available in X-Y mode and Roll mode), press MENU HorRef in the horizontal control area (HORIZONTAL) to select the desired reference mode and the default is “Center”. 1.
3 To Set the Horizontal System RIGOL horizontally around the user-defined reference position. In the horizontal direction, the screen can display a maximum of 700 points with the leftmost as 350 and the rightmost as -350. For example, set the reference position to 150.
4 To Set the Sample System RIGOL 4 To Set the Sample System The contents of this chapter: Acquisition Mode Sample Mode Sample Rate Memory Depth Antialiasing DS4000 User’s Guide 4-1
4 To Set the Sample System RIGOL Acquisition Mode The acquisition mode is used to control how to generate waveform points from sample points. Press Acquire Acquisition in the function menu at the front panel and use to select the desired acquisition mode (the default is normal), then press down the knob to select this mode. You can also press Acquisition continuously to switch the acquisition mode.
4 To Set the Sample System When “Average” mode is selected, press Averages and use number of averages as the power function of 2.
RIGOL 4 To Set the Sample System Peak Detect In this mode, the oscilloscope acquires the maximum and minimum values of the signal within the sample interval to get the envelope or the narrow pulse of the signal that might be lost. In this mode, signal confusion can be prevented but the noise displayed would be larger. In this mode, the oscilloscope can display all the pulses with pulse widths at least as wide as the sample period.
4 To Set the Sample System RIGOL Sample Mode This oscilloscope only supports real-time sample. In this mode, the oscilloscope samples and displays waveform within a trigger event. The maximum real-time sample rate of DS4000 is 4 GSa/s and the current sample rate is displayed under the Sa Rate menu. Tip Press RUN/STOP to stop the sample, the oscilloscope will hold the last display. At this point, you can still use the vertical control and horizontal control to pan and zoom the waveform.
RIGOL 4 To Set the Sample System Sample Rate Sample refers to the process that the oscilloscope converts analog signals to digital signals at a certain time interval and stores the data in order. Sample rate is the reciprocal of the time interval. Sample Rate 1 /Δ t The sample rate of this oscilloscope is up to 4 GSa/s.
4 To Set the Sample System 3. RIGOL Waveform Leakage: when the sample rate is too low, the waveform rebuilt from the sample data does not reflect all the actual signal information.
4 To Set the Sample System RIGOL Memory Depth Memory depth refers to the number of waveform points that the oscilloscope can store in a single trigger sample and it reflects the storage ability of the sample memory. DS4000 provides up to 140 M points memory depth.
4 To Set the Sample System RIGOL Antialiasing At slower sweep speed, the sample rate is reduced and a dedicated display algorithm is used to minimize the possibility of aliasing. Press Acquire Anti_aliasing to enable or disable the antialiasing function. By default, antialiasing is disabled. The displayed waveforms will be more susceptible to aliasing when this function is disabled.
5 To Trigger the Oscilloscope RIGOL 5 To Trigger the Oscilloscope During work, the oscilloscope samples waveform data continuously no metter whether the trigger is stable, but only stable trigger can garrantee stable display. The trigger circuit guarantees that each sweep or sample starts from the trigger condition defined by users on the input signal, namely each sweep and sample are synchronous and the waveforms acquired overlap to display stable waveform.
5 To Trigger the Oscilloscope RIGOL Trigger Source Press MENU Source in the trigger control area (TRIGGER) at the front panel to select the desired trigger source. Siganls input from CH1 to CH4 and the [EXT TRIG] connector as well as the AC Line can all be used as trigger source. MENU Analog channel input: Signals input from analog channels CH1 to CH4 can all be used as the trigger source. No matter whether the input of the channel selected is enabled, the channel can work normally.
5 To Trigger the Oscilloscope RIGOL Trigger Mode Trigger mode affects the way in which the oscilloscope searches for the trigger. The following is the schematic diagram of the acquisition memory. As shown in the figure below, the position of the trigger event in the acquisition memory is determined by the reference time point and the delay setting. Trigger Event Pre-trigger Buffer Post-trigger Buffer Acquisition memory Pre-trigger/Delayed trigger: Acquire data before/after the trigger event.
RIGOL 5 To Trigger the Oscilloscope Auto: No matter whether the trigger condition is met, there is always waveform display. A horizontal line is displayed when no signal is input. In this mode, the oscilloscope operates by first filling the pre-trigger buffer. It starts searching for a trigger after the pre-trigger buffer is filled and continues to flow data through this buffer while it searches for the trigger.
5 To Trigger the Oscilloscope RIGOL Trigger Coupling Trigger coupling decides which kind of components will be transmitted to the trigger circuit. Please distinguish it with “Channel Coupling”. DC: allow DC and AC components into the trigger path. AC: block all the DC components and attenuate signals lower than 8 Hz. LF reject: block the DC components and reject the low frequency components (lower than 5 kHz). HF reject: reject the high frequency components (higher than 50 kHz).
5 To Trigger the Oscilloscope RIGOL Trigger Holdoff Trigger holdoff can be used to stabilize the display of complex waveforms (such as pulse series). Holdoff time is the amount of time that the oscilloscope waits before re-arming the trigger circuitry. The oscilloscope will not trigger until the holdoff time expires.
5 To Trigger the Oscilloscope RIGOL Noise Rejection Noise Rejection adds additional hysteresis to the trigger circuitry. Noise rejection reduces the possibility of noise trigger but also reduces the trigger sensitivity; therefore, a greater amplitude waveform is required to trigger the oscilloscope. Press MENU Setting Noise Reject in the trigger control area (TRIGGER) at the front panel to enable or disable noise rejection.
RIGOL 5 To Trigger the Oscilloscope Trigger Type DS4000 provides various trigger functions, including various serial bus triggers.
5 To Trigger the Oscilloscope RIGOL Edge Trigger Trigger on the trigger threshold of the specified edge of the input signal. Trigger Type: Press Type to select “Edge”. At this point, the trigger setting information as shown in the figure below is displayed at the upper right corner of the screen. Source Selection: Press Source to select CH1 to CH4, EXT, EXT/5 or AC Line as the Trigger Source. The current trigger source is displayed at the upper right corner of the screen.
RIGOL 5 To Trigger the Oscilloscope turning the knob, the trigger level line and the trigger mark disappear in about 2 seconds.
5 To Trigger the Oscilloscope RIGOL Pulse Trigger Trigger on the positive or negative pulse with a specified width. Trigger Type: Press Type to select “Pulse”. At this point, the trigger setting information as shown in the figure below is displayed at the upper right corner of the screen. Source Selection: Press Source to select CH1 to CH4 or EXT as the Trigger Source. The current trigger source is displayed at the upper right corner of the screen.
5 To Trigger the Oscilloscope RIGOL A B Trigger level Positive pulse width When the Pulse Condition is set to , , or , press Setting and use to input the desired value. The range available is from 4 ns to 4 s. When the Pulse Condition is set to or , press Upper Limit and Lower Limit and use to input the desired values respectively. The range of the upper limit is from 12 ns to 4 s. The range of the lower limit is from 4 ns to 3.99 s.
5 To Trigger the Oscilloscope RIGOL Slope Trigger Trigger on the positive or negative slope of specified time. Trigger Type: Press Type to select “Slope”. At this point, the trigger setting information as shown in the figure below is displayed at the upper right corner of the screen. Source Selection: Press Source to select CH1 to CH4 as the Trigger Source. The current trigger source is displayed at the upper right corner of the screen.
5 To Trigger the Oscilloscope RIGOL A B The upper limit of trigger level (Up Lvl) The lower limit of trigger level (Low Lvl) Positive Slope time When the Slope Condition is set to , , or , press Time and use to input the desired value. The range available is from 10 ns to 1 s. When the Slope Condition is set to or , press Upper Limit and Lower Limit and use to input the desired values respectively. The range of time upper limit is from 20 ns to 1 s.
5 To Trigger the Oscilloscope RIGOL “Low Lvl” and “Slew Rate” change accordingly but “UP Lvl” remains unchanged. : adjust the upper and lower limits of the trigger level at the same time. During the adjustment, “UP Lvl” and “Low Lvl” change accordingly but “Slew Rate” remains unchanged. Trigger Mode: Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto, normal or single. The corresponding status light of the current trigger mode turns on.
RIGOL 5 To Trigger the Oscilloscope Video Trigger Trigger on the standard video signal field or line of NTSC (National Television Standards Committee), PAL (Phase Alternating Line), SECAM (sequential color with memory) or HDTV (High Definition Television). Trigger Type: Press Type to select “Video”. At this point, the trigger setting information as shown in the figure below is displayed at the upper right corner of the screen. Source Selection: Press Source to select CH1 to CH4 as the Trigger Source.
5 To Trigger the Oscilloscope RIGOL Video Standard: Press Standard to select the desired video standard. NTSC: the field frequency is 60 fields per second and the frame frequency is 30 frames per second. The TV sweep line is 525 with the even field goes first and the odd field follows behind. PAL: the frame frequency is 25 frames per second. The TV sweep line is 625 with the odd field goes first and the even field follows behind. SECAM: the frame frequency is 25 frames per second.
RIGOL 5 To Trigger the Oscilloscope Pattern Trigger Identify a trigger condition by looking for a specified pattern. This pattern is a logical “AND” combination of the channels. Each channel can have a value of high (H), low (L) or don’t care (X). A rising or falling edge can be specified for one channel included in the pattern. When an edge is specified, the oscilloscope will trigger at the edge specified if the pattern set for the other channels are true (H or L).
5 To Trigger the Oscilloscope RIGOL pattern, the former edge item defined will be replaced by X. Trigger Mode: Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto, normal or single. The corresponding status light of the current trigger mode turns on. Trigger Setting: Press Setting to set the trigger parameters (trigger coupling, trigger holdoff and noise rejection) under this trigger type. Trigger Level: Use TRIGGER LEVEL to modify the level.
RIGOL 5 To Trigger the Oscilloscope RS232 Trigger Trigger according to the start frame, error frame, check error or data. Below is the explanatory figure of RS232 protocol. Trigger Type: Press Type to select “RS232”. At this point, the trigger setting information as shown in the figure below is displayed at the upper right corner of the screen. Source Selection: Press Source to select CH1 to CH4 as the Trigger Source. The current trigger source is displayed at the upper right corner of the screen.
5 To Trigger the Oscilloscope RIGOL the upper limits are 31, 63, 127 and 255 respectively; Baud Rate: Set the baud rate of data transmmision (equal to specifying a clock frequency). Press Baud to set the desired baud rate to 2400 bps, 4800 bps, 9600 bps (default), 19200 bps, 38400 bps, 57600 bps, 115200 bps and user-defined. When “User” is selected, press Setup and use to set a more specific value from 1 to 900000 with a adjustment step of 1 bps.
RIGOL 5 To Trigger the Oscilloscope I2C Trigger Trigger on the start condition, restart, stop, missing acknowledgement or on the read/write frame with specific device address and data value. In I2C trigger, you need to specify the SCL and SDA data sources. The figure below shows the complete data transmission of I2C bus. Trigger Type: Press Type to select “I2C”. At this point, the trigger setting information as shown in the figure below is displayed at the upper right corner of the screen.
5 To Trigger the Oscilloscope RIGOL Address: trigger on the clock (SCL) edge corresponding to the byte of data (SDA) behind the preset address (Write, Read or R/W direction). After this trigger condition is selected: --press AddrBits to select “7 bit” or “10 bit”; --press Address to set the address value according to the setting in AddrBits and the ranges are from 0 to 127 and from 0 to 1023 respectively; --press Direction to select “Read”, “Write” or “R/W”.
RIGOL 5 To Trigger the Oscilloscope Trigger Setting: Press Setting to set the trigger parameters (trigger coupling, trigger holdoff and noise rejection) under this trigger type. Trigger Level: Use TRIGGER LEVEL to modify the trigger level of SCL or SDA channel. For details, please refer to the description of “Trigger Level” on page 5-9.
5 To Trigger the Oscilloscope RIGOL SPI Trigger Trigger on the data pattern on the specified edge. When using SPI trigger, you need to specify the SCL, SDA and CS data sources. Below is the sequence chart of SPI bus data transmission. Trigger Type: Press Type to select “SPI”. At this point, the trigger setting information as shown in the figure below is displayed at the upper right corner of the screen. Source Selection: Press SCL, SDA and CS to specify the data sources of SCL, SDA and CS respectively.
5 To Trigger the Oscilloscope RIGOL this condition is selected. TimeOut: set the minimum time that the clock (SCL) signal must be idle before the oscilloscope starts to search for the data (SDA) on which to trigger. After this trigger condition is selected, press TimeOut to set the timeout value and the range is from 100 ns to 1 s. Note that, at this point, CS is invalid (not displayed).
5 To Trigger the Oscilloscope RIGOL CAN Trigger Trigger on the specified frame type of the data frame. When using CAN trigger, you need to specify the signal source, signal rate and trigger signal type of the CAN signal. The figure below shows the standard and expanded formats of CAN bus data frame. Trigger Type: Press Type to select “CAN”. At this point, the trigger setting information as shown in the figure below is displayed at the upper right corner of the screen.
RIGOL 5 To Trigger the Oscilloscope CAN_H: actual CAN_H bus signal. CAN_L: actual CAN_L bus signal. Differential: CAN differential bus signal connected to the analog channel via the differential probe. Trigger Condition: Press When to select the desired trigger condition. SOF: trigger on the start frame of the data frame. EOF: trigger on the end frame of the data frame. Frame Type: after this type is selected, press Frame Type to select to trigger on “Data”, “Remote”, “Error” or “OverLoad”.
5 To Trigger the Oscilloscope press User and use RIGOL to input the desired rate within 1 kb/s and 10.0 Mb/s. Sample Point: Sample point is the point within the bit’s time. The oscilloscope samples the bit level at this point. “Sample point” is represented by the percentage of “the time from the start of the bit’s time to the sample point time” in the “bit’s time”. Press SamplePoint and use to modify the parameter with a step of 1% and the range is from 5% to 95%.
RIGOL 5 To Trigger the Oscilloscope FlexRay Trigger Trigger on the specified frame, symbol, error or TSS (Transmission Start Sequence) of FlexRay bus. FlexRay is a kind of differential serial bus configured with three continuous segments (namely packet header, payload and packet end). Its data transmission rate is up to 10 Mbps. Each frame contains a static and dynamic segment and ends with the bus idle time.
5 To Trigger the Oscilloscope RIGOL Trigger Condition: Press When to select the desired trigger condition. Frame: trigger on the frame of FlexRay bus. Symbol: trigger on the CID (Channel Idle Delimiter), CAS (Collision Avoidance Symbol), MTS (Media Access Test Symbol) and WUP (Wakeup Pattern) of FlexRay bus. Error: trigger when error occurs to FlexRay bus, including header CRC error and frame CRC error. TSS: trigger on the transmission start sequence of FlexRay bus.
RIGOL 5 To Trigger the Oscilloscope USB Trigger Trigger on the SOP, EOP, RC, Suspend and Exit Suspend of the data packet on the differential USB data cable (D+ and D-). This trigger supports USB Low Speed and Full Speed. The figure below shows the USB data transmission protocol. Trigger Type: Press Type to select “USB”. At this point, the trigger setting information as shown in the figure below is displayed at the upper right corner of the screen.
5 To Trigger the Oscilloscope RIGOL Trigger Setting: Press Setting to set the trigger parameters (trigger coupling, trigger holdoff and noise rejection) under this trigger type. Trigger Level: Use TRIGGER LEVEL to modify the level. For details, please refer to the description of “Trigger Level” on page 5-9.
RIGOL 5 To Trigger the Oscilloscope Trigger Output Connector The trigger output connector at the rear panel can output trigger signals determined by the current setting. Trigger Output Connector Press Utility Aux Output to select “TrigOut”. When the oscilloscope is triggered, it will output a trigger signal determined by the current trigger setting through the [Trig Out/Calibration] connector.
6 To Make Measurements RIGOL 6 To Make Measurements DS4000 can make math operation, cursor measurement and auto measurement on sampled and displayed data.
6 To Make Measurements RIGOL Math Operation DS4000 can realize various math operations (including: addition (A+B), subtraction (A-B), multiplication (AxB), division (A÷ B), FFT, logic operation and advanced operation) of waveforms between channels. The results of math operation also allows further measurement (for details, please refer to “Cursor Measurement”). MATH Press MATH Operate in the vertical control area (VERTICAL) at the front panel to select the desired operation function.
6 To Make Measurements RIGOL Substraction Subtract the waveform voltage values of signal source B from that of source A point by point and display the results. Press MATH Operate to select “A-B”: Press Source A and Source B to select the desired channels. The channels available are CH1, CH2, CH3 and CH4. Press Invert to turn the inverted display of the operation results on or off. Press and use to adjust the vertical position of the operation results.
6 To Make Measurements RIGOL Division Divide the waveform voltage values of signal source A by that of source B point by point and display the results. It can be used to analyze the multiple relationships of waveforms in two channels. Note that when the voltage value of channel B is 0, the result of the division is treated as 0. Press MATH Operate to select “A ÷ B”: Press Source A and Source B to select the desired channels. The channels available are CH1, CH2, CH3 and CH4.
6 To Make Measurements RIGOL FFT FFT is used to quickly perform Fourier transform on specified signals and transform time domain signals to frequency domain signals. FFT operation can facilitate the following works: Measure harmonic components and distortion in the system Measure the characteristics of the noise in DC power Analyze vibration Press MATH Operate to select “FFT” and set the parameters of FFT operation.
6 To Make Measurements RIGOL characteristics and are applicable to measure different waveforms. You need to select the window function according to different waveforms and their characteristics. Press Window to select the desired window function and the default is “Rectangle”. Table 6-1 Window Functions 3.
6 To Make Measurements HORIZONTAL 4. RIGOL SCALE to switch between “Split” and “Full Screen”. Set the Vertical Scale In FFT measurement, the unit of the horizontal axis changes from time to frequency. Use HORIZONTAL SCALE and HORIZONTAL POSITION to set the scale and position of the horizontal axis respectively. The unit of the vertical axis can be dBVrms or Vrms which use logarithmic mode and linear mode to display vertical amplitude respectively.
6 To Make Measurements RIGOL Logic Operation Perform logic operation on the waveform voltage values of the specified sources point by point and display the results. In operation, when the voltage value of the source channel is greater than 0, it is regarded as logic “1”; otherwise logic “0”.
6 To Make Measurements RIGOL Press MATH Operate to select “Logic”: Press Expression to select the desired operation expression and the default is “AND”. Press Source A and Source B to select the desired channels. The channels available are CH1, CH2, CH3 and CH4. Press Invert to turn the inverted display of the operation results on or off. Press and use Press and use to adjust the vertical scale of the operation results.
RIGOL 6 To Make Measurements Advanced Operation DS4000 provides advanced operation function that allows users to define operation functions. Press MATH Operate Advance Expression ON and the editing window as shown in the figure below is displayed. Turn to select any item in “Channel”, “Function”, “Variable”, “Operator” and “Figure” (if they are currently available for selection), then press down the knob and the item selected will de displayed in the entry box on the right of the “Expression”.
6 To Make Measurements 3. RIGOL Function Please refer to the following table to get the functions of each function. Note that the left brackets “(” here are only used to facilitate your entry and they are not a part of the function names. Table 6-3 Functions 4. Name Function Intg( Calculate the integral of the selected source. You can use integral to measure the area under a waveform or the pulse energy. Diff( Calculate the discrete time differentiate of the selected source.
6 To Make Measurements RIGOL For example, Variable1 is set to 6.1074× 108 via the following settings. Variable: Variable1 Mantissa: 6.1074 Exponent: 8 5. Operator Please refer to the following table to get the functions of each operator.
6 To Make Measurements RIGOL Auto Measurement DS4000 provides auto measurements of 22 waveform parameters and the statistics and analysis of the measurement results. What’s more, you can also use the frequency counter to realize more precise frequency measurement.
RIGOL 6 To Make Measurements should be no lower than 50 Hz, the duty cycle be greater than 1% and the amplitude be at least 20 mVpp. If the parameters of the signal under test exceed these limits, after this softkey is pressed, the quick parameter measurement items might not be displayed in the pop-up menu.
6 To Make Measurements RIGOL One-key Measurement of 22 Parameters Press MENU at the left of the screen to turn on the measurement menu of the 22 parameters and then press the corresponding menu softkey to quickly realize “One-key” measurement. The measurement result will be displayed at the bottom of the screen. The icons of time and voltage parameters in the measurement items and the measurement results on the screen are always marked in the same color with the channel (Measure Source) currently used.
6 To Make Measurements RIGOL Time Parameters RiseTime FallTime 90% 50% 10% +Width 1. 2. 3. 4. 5. 6. 7. 8. 6-16 -Width Period: defined as the time between the middle threshold points of two consecutive, like-polarity edges. Frequency: defined as the reciprocal of period. Rise Time: the time for the signal amplitude to rise from 10% to 90%. Fall Time: the time for the signal amplitude to fall from 90% to 10%.
6 To Make Measurements RIGOL Delay and Phase Period Source A Delay Source B Source A and source B can be any channel from CH1 to CH4. 1. Delay A→B : the time difference between the rising edges of source A and 2. source B. Negative delay indicates that the selected rising edge of source A occurred after the selected edge of source B. Delay A→B : the time difference between the falling edges of source A and 3. source B.
6 To Make Measurements RIGOL Voltage Parameters Overshoot Vmax Vtop Vamp Vpp Vbase Vmin Preshoot 1. 2. 3. 4. 5. 6. 7. Vmax: the voltage value from the highest point of the waveform to the GND. Vmin: the voltage value from the lowest point of the waveform to the GND. Vpp: the voltage value from the highest point to the lowest point of the waveform. Vtop: the voltage value from the flat top of the waveform to the GND. Vbase: the voltage value from the flat base of the waveform to the GND.
6 To Make Measurements RIGOL n the gating area. RMS x i1 n 2 i , where, x i is the ith point being measured, n is the number of points being measured. 9. Overshoot: the ratio of the difference of the maximum value and top value of the waveform to the amplitude value. 10. Preshoot: the ratio of the difference of the minimum value and base value of the waveform to the amplitude value.
RIGOL 6 To Make Measurements Frequency Counter Measurement The hardware frequency counter supplied with this oscilloscope can make more precise measurement of the input signal frequency. Press Measure Counter to select anyone of CH1 to CH4 as the measurement source. The measurement result is displayed at the upper right corner of the screen and you can identify the current measurement source according to the color of the icon.
6 To Make Measurements RIGOL Measurement Setting 1. Source Selection Press Measure Source to select the desired channel for measurement (CH1 to CH4). The color of the parameter icons under MENU at the left of the screen will change with the source selected. 2. Measurement Range Press Measure Meas.Range Meas.Range to select “Screen Region” or “Cursor Region” for measurement. When “Cursor Region” is selected, two cursor lines appear on the screen.
RIGOL 6 To Make Measurements and “Phase A→B ”. Press Measure Meas.Setting Type “Phase” and then press Source A and Source B to set the two channel sources (CH1 to CH4) of phase measurement respectively. 5. Threshold Measurement Setting Specify the vertical level (in percentage) being measured in the analog channel. Measurements of all the time, delay and phase parameters will be influenced by this setting. Press Measure Meas.
6 To Make Measurements RIGOL To Clear the Measurement If you have currently enabled one or more items in the 22 measurement parameters, you can “Delete” or “Recover” the first five parameters or “Delete” or “Recover” all the measurement items enabled. Note that the first five parameters are determined according to the order in which you turned them on and they will not change as you delete one or more measurement items.
RIGOL 6 To Make Measurements All Measurement All measurement could measure all the time and voltage parameters (each measurement source has 18 items, measurements can be performed on the four measurement sources at the same time) of the current measurement source and display the results atthe screen. Press Measure All Measure to enable or disable the all measurement function. Press All Measure Source and use to select the channel(s) to be measured (CH1 to CH4).
6 To Make Measurements RIGOL Statistic Function Make statistic and display the current, average, minimum (or standard deviation) and maximum (or count) values of at most 5 measurement items that are turned on last. Press Measure Statistic to turn the statistic function on or off. Press Measure StatisSel to select “Extremum” or “Difference” measurement. When “Extremum” is selected, minimum and maximum values are displayed.
RIGOL 6 To Make Measurements Measurement History To view the history measurement data, press Measure MeasHistory MeasHistory “ON”. The history data can be displayed in two modes: Graph: display the results of the last 10 measurements of at most 5 measurement items that are turned on last in graph mode. The measurement points are connected using linear interpolation. Table: display the results of the last 10 measurements of at most 5 measurement items that are turned on last in table mode.
6 To Make Measurements RIGOL Cursor Measurement Cursors are the horizontal and vertical marks that can be used to measure the X axis values (usually Time) and Y axis values (usually Voltage) on a selected waveform. Please connect the signal to the oscilloscope and obtain stable display before using cursor measurement. All the “Auto Measurement” parameters can be measured through cursor measurement.
RIGOL 6 To Make Measurements Manual Mode In this mode, a pair of cursors will appear. You can adjust the cursors manually to measure the X (or Y), X increment (or Y increment) between cursors and the reciprocal of X increment on the waveform of the selected source (CH1 to CH4 or MATH). Press Cousor Mode “Manual” to turn the manual cursor function on and the measurement results will be displayed at the upper left corner of the screen in the following mode.
6 To Make Measurements 3. RIGOL Select X (Y) Axis Unit When the cursor type is “Time”, press Time Unit to select “s”, “Hz”, “°” or “%”. s: when this unit is selected, in the measurement results, CurA, CurB and △X are in “s” and 1/△X is in “Hz”. Hz: when this unit is selected, in the measurement results, CurA, CurB and △X are in “Hz” and 1/△X is in “s”. °: when this unit is selected, in the measurement results, CurA, CurB and △X are in “°”.
6 To Make Measurements RIGOL 5. Measurement Example Use manual cursor measurement to measure the period (△X) of a square waveform and the result is 1 ms equaling the result from auto measurement.
6 To Make Measurements RIGOL Track Mode In this mode, one or two pairs of cursors will appear. You can adjust the two pairs of cursors to measure the X and Y values on two different sources respectively. The points being measured on cursor A and B are marked by an orange rectangle and rhombus respectively. When the cursors are moved horizontally, the marks will position on the waveform automatically.
RIGOL 6 To Make Measurements If needed, please refer to the following steps to modify track cursor measurement parameters. 1. Select Measurement Source Press Cursor A to select the waveform of analog channels (CH1 to CH4) or math operation results (MATH) as the measurement source of cursor A (only channels enabled are available). You can also select “None”, namely do not use cursor A.
6 To Make Measurements RIGOL Cursor Track (Before Horizontal Expansion): Cursor Track (After Horizontal Expansion): DS4000 User’s Guide 6-33
RIGOL 6 To Make Measurements Auto Mode In this mode, one or more cursors will appear. You can use auto cursor measurement to measure any of the 22 waveform parameters. Before using this mode, you need to at least enable one auto measurement parameter and the number of cursors will change with the measurement parameter enabled.
7 Protocol Decoding RIGOL 7 Protocol Decoding Protocol analysis can be used by users to discover errors, debug hardware and accelerate development easily, so as to guarantee quick and high-quality accomplishment of projects. Protocol decoding is the basis of protocol analysis. Only protocol analyses with correct protocol decoding are acceptable and only correct protocol decoding can provide more error information.
7 Protocol Decoding RIGOL Parallel Decoding Parallel bus consists of clock line and data line. As shown in the figure below, CLK is the clock line, while Bit0 and Bit1 are the 0 bit and 1st bit on the data line respectively. CLK Bit0 Bit1 The oscilloscope will sample the channel data on the rising edge, falling edge or the rising&falling edges of the clock and judge each data point (logic “1” or logic “0”) according to the preset threshold level.
7 Protocol Decoding RIGOL example, when the bus bits is 20, the range available is 0, 1…19. Next, press Channel to specify a channel source for the bit currently selected in CurrentBit. 3. Analog Channel Threshold Setting To judge logic “1” and logic “0” of the buses, you need to set a threshold for each analog channel (CH1, CH2, CH3 and CH4). When the signal amplitude is greater than the preset value, it is considered as “1”; otherwise “0”.
RIGOL 5. 7 Protocol Decoding Decoding Table The decoding table displays the decoded data and the corresponding time in table format. It can be used to observe relatively longer decoded data to solve the problem that some data could not be seen clearly on the screen. Press Event Table Event Table to select “ON” (note that this operation is only available when BusStatus is set to “ON”) to enter the decoding table interface as shown in the figure below.
7 Protocol Decoding RIGOL RS232 Decoding (Option) RS232 serial bus consists of the transmitting data line (TX) and the receiving data line (RX). Rx Tx DeviceA DeviceB Tx Rx The industry standard of RS232 uses “Negative Logic”, namely high level is logic “0” and low level is logic “1”. 0 1 0 1 1 0 1 0 1 1 By default, RS232 uses LSB (Least Significant Bit) transmission sequence, namely the lowest bit of the data is transmitted first.
7 Protocol Decoding RIGOL Stop Bit Check Bit Start Bit Data Bits Start Bit: represent when the data starts outputting. Setting the Polarity is equivalent to specifying the “Start Bit”. Data Bits: represent the number of data bits actually contained in each frame of data. Even-OddCheck: be used to check the correctness of the data transmission. Odd-Check: the number of “1” in the data bit and check bit is an odd.
7 Protocol Decoding RIGOL 5. Data Packet Setting As mentioned before, in RS232, you need to set the start bit, data bits, check bit (optional) and stop bit of each frame of data. “Start Bit” is specified by the “Polarity Setting”. The setting methods of other parameters are as follows: Press Data Bits to set the data width of each frame. It can be set to 5, 6, 7, 8 or 9 and the default is 8. Press Stop Bit to set the stop bit after each frame of data. It can be set to 1 bit, 1.5 bits or 2 bits.
7 Protocol Decoding RIGOL 8. The Error Expression during Decoding DS4000 makes full use of the resources such as color and view to express the results of the protocol decoding effectively so as to let users find the desired information quickly. End Frame Error: Errors generated when the end frame condition is not met. When the stop bit is set to 1.
7 Protocol Decoding RIGOL Check Error: When check bit error is detected during the decoding, red error mark will be displayed. For example, when the transmitting terminal is set to none check and the decoder is set to odd check, the following check error occurs: (00100000,LSB) The check bit detected is 1 Wherein, there are odd number (1) of 1 in the 8 bits data 00100000 and the check bit should be 0; but the check bit detected on the TX is 1, thus check error occurs.
7 Protocol Decoding RIGOL SPI Decoding (Option) SPI serial bus consists of chip select line (SS), clock line (SCLK), MISO and MOSI. SS SS Host SS MISO MISO MOSI MOSI SCLK SCLK SCLK Device MISO MOSI SS: set the polarity to “Low” or “High”. When SS selection is valid, the SPI bus samples data from MISO and MOSI at the jumping point of SCLK. SCLK: sample data from MISO and MOSI on the clock rising edge or falling edge. MISO: master input/slave output. Set the polarity to “Low” or “High”.
7 Protocol Decoding RIGOL 3. MISO Setting Press MISO to enter the MISO data line setting interface. Press Channel to select any channel (CH1 to CH4) as the MISO data channel. When “OFF” is selected, this data line is not set. Press Polarity to set the polarity of the MISO data line to “Low” or “High”. Press Threshold to set the threshold of the MISO data channel. 4. MOSI Setting Press MOSI to enter the MOSI data line setting interface.
RIGOL 7 Protocol Decoding external USB storage device in CSV format. 9. Error Expressions during Decoding When the clock for a frame in SPI is nout enough, the data is filled with red patches. For example, when Data Bits is set to 7 and SCLK slope is set to rising edge, decoding error will be generated.
7 Protocol Decoding RIGOL I2C Decoding (Option) I2C serial bus consists of the clock line (SCLK) and the data line (SDA). Vcc A2 SCLK SDA Device A1 A0 Host SCLK SDA SCLK: sample the SDA on the clock rising edge or falling edge. SDA: denote the data channel. Press Decode1 Decode to select “I2C” and open the I2C decoding function menu. 1. SCLK Setting Press SCLK to select any channel (CH1 to CH4) as the clock channel. Press SCLKThreshold to set the threshold of the clock channel. 2.
RIGOL 7 Protocol Decoding Press BusStatus to turn the bus display on or off. 4. Decoding Table The decoding table displays the decoded data, the correspondint time, data direction, ID and ACK information in table format. Press Event Table Event Table to select “ON” (note that this operation is only available when BusStatus is set to “ON”) to enter the decoding table interface as shown in the figure below.
7 Protocol Decoding RIGOL When the ACK (ACKnowledge Character) is not met, the red error marks as shown in the figure below will be displayed.
7 Protocol Decoding RIGOL CAN Decoding (Option) Press Decode1 Decode and select “CAN” to open the CAN decoding function menu. 1. Source Press Source and select any channel (CH1 to CH4) as the source channel. 2. Signal Type Press Signal Type to select the desired signal type. CAN_H: the actual CAN_H bus signal. CAN_L: the actual CAN_L bus signal. Differential: the CAN differential bus signals connected to an analog channel using a differential probe.
7 Protocol Decoding RIGOL 5. Threshold Refer to the introduction in “Parallel Decoding”. 6. Display-related Setting Press Format to set the bus display format to Hex, Decimal, Binary or ASCII. Press Offset and use to adjust the vertical display position of the bus. Press BusStatus to enable or disable bus display. 7. Decoding Table The decoding table displays the decoded data, the corresponding time, frame ID, DLC, CRC and ACK information in table format.
7 Protocol Decoding RIGOL and red error mark is displayed when error occurs.
7 Protocol Decoding RIGOL FlexRay Decoding (Option) Press Decode1 Decode and select “FlexRay” to open the FlexRay decoding function menu. 1. Source Press Source to select any channel (CH1 to CH4) as the signal source channel. 2. Signal Path Press Signal Path to select the signal path (A or B) that matches the FlexRay bus signal. 3. Signal Type Press Signal Type to select the type of signal that matches the FlexRay bus. The signal types available include BP, BM and TX/RX. 4.
RIGOL 7 Protocol Decoding 7. Display-related Setting Press Package to set whether to display the FlexRay in package. Press Format to set the display format of the bus to Hex, Decimal, Binary or ASCII. Press Offset and use to adjust the vertical display position of the bus. Press BusStatus to enable or disable bus display. 8. Decoding Table The decoding table lists the decoded data, the corresponding time and error information in table format.
7 Protocol Decoding RIGOL TSS: transmission start sequence and is expressed by light purple patch. Frame Type: FlexRay frame can be NORMAL, SYNC, SUP or NULL. The frame type in the figure above is “NOR” and is expressed by purple patch. Frame ID: decimal number and is expressed by blue patch. Effective Loading Length: decimal number and is expressed by brown patch. Head CRC: hexadecimal number and is expressed by blue-green patch. When CRC is invalid, it is expressed by red patch.
RIGOL 7 Protocol Decoding End CRC: hexadecimal number and is expressed by blue-green patch. When CRC is invalid, it is expressed by red patch. DTS: dynamic end sequence and is expressed by light purple patch.
8 Reference Waveform RIGOL 8 Reference Waveform In actual testing process, the waveform being tested can be compared with the reference waveform to judge the causes of failures.
8 Reference Waveform RIGOL To Enable REF Function Press REF in the vertical control area (VERTICAL) at the front panel to enable the REF function. Note that when the time base is in X-Y mode, REF function can not be enabled. REF DS4000 provides 10 reference waveform channels. Press Channel and use set the desired reference channel to on or off and a channel icon (for example, of the channel enabled will be display at the left side of the screen grid.
8 Reference Waveform RIGOL To Select REF Source Press Current and use to select any reference channel (Ref1 to Ref10) that has been turned on and then press Source to specify a reference source (CH1 to CH4, MATH) for this channel. To Save to Internal Memory Press Save to save the waveform (screen region) in the specified source to internal memory as reference waveform and display it on the screen.
RIGOL 8 Reference Waveform To Export to Internal or External Memory Users can also save the reference waveform to the internal Flash memory or external USB storage device. The file format of the reference waveform is “*.ref”. At most 10 reference files (LocalREF0.ref to LocalREF9.ref) can be saved inside the instrument. Press Export to enter file store interface. Please refer to the relative descriptions in “Store and Recall” to save the reference waveform to internal or external memory.
9 Pass/Fail Test RIGOL 9 Pass/Fail Test Monitor the change of the signal by judging whether the input signal is within the mask created. The test results can be displayed on the screen as well as be declared by the system sound or the pulse signal output from the [Trig Out/Calibration] connector at the rear panel.
9 Pass/Fail Test RIGOL To Enable Pass/Fail Test Press Utility Pass/Fail Enable Test to select “ON”. Note that when the time base is in X-Y mode, the Pass/Fail test function can not be enabled. Utility To start testing, press Enable Test and select “ON”. Then, press Operate to select “ ” to start testing and select “■” to stop testing. The figure below is the test interface: You can select the source, create mask as well as save and load the test mask.
9 Pass/Fail Test RIGOL To Select Source Press Source to select the channel (CH1 to CH4) to be tested. During the test, the oscilloscope will judge whether each frame of waveform in the source complies with the current test mask and those waveforms pass through the mask area (blue area) is considered as failed. To Create Mask Users can define their own test masks. Press Enable Test to turn the mask display area on. Press X Mask and Y Mask respectively and turn , the mask lines appear on the screen.
9 Pass/Fail Test RIGOL Test and Ouput Before the test, you can use the following method to set the output mode of the test results. Press Msg Display to select “ON” or “OFF”. When “ON” is selected, the test results will be displayed at the upper right corner of the screen. Press Stop On Outp to select “ON” or “OFF”. ON: when failed waveforms are detected, the oscilloscope will stop the test and enter the “STOP” state.
9 Pass/Fail Test RIGOL To Save the Test Mask Users can save the current test mask to the internal Flash memory or external USB storage device. The file format of the test mask file is “*.pf”. The internal memory can store at most 10 test mask files (LocalPF.pf). Press Save to enter file store interface. Please refer to the relative descriptions in “Store and Recall” to save the test mask file to the internal or external memory. To Load the Test Mask Users can also load the test mask files (*.
10 Waveform Record RIGOL 10 Waveform Record Waveform record can record the waveforms of the input channels (CH1 to CH4). In record constant on mode, the oscilloscope can record the input waveform contimuously until users press RUN/STOP. Waveform playback and analysis can provide better waveform analysis effect. (Note that the horizontal time base must be set to Y-T mode during waveform record.
10 Waveform Record RIGOL Waveform Record Waveforms from all the channels currently turned on will be recorded during waveform record. Navigation Knob (Playback) Waveform Record Play/Pause (Playback) Stop Recording (Playback)/OFF Press Record Mode and use operation menu. 1. to select “Record” to open the waveform record 1. End Frame Press End Frame and use to set the desired number of frames to be recorded. The number of frames available is related to the memory depth currently selected. 2.
10 Waveform Record 4. RIGOL Maximum Number of Frames The menu shows the maximum number of frames that can be recorded currently. As the capacity of the waveform memory is fixed, the more the number of points each frame of waveform has, the less the number of waveform frames can be recorded. Thus, the maximum end frame of waveform record is decided by the “Memory Depth” currently selected. Please refer to the instruction in “Memory Depth” to select the desired memory depth.
RIGOL 10 Waveform Record Record Constant On DS4000 provides record “Open” mode for waveform record. When this mode is enabled, the oscilloscope records the waveform continuously according to the current memory depth. The length of waveform recorded is limited by the memory depth and the waveform data recorded before will be overwritten by the waveform currently recorded.
10 Waveform Record RIGOL 3. Measure the signal parameters in record constant on mode. 4. Press RUN/STOP to stop the record. 5. Analyze the waveform recorded using waveform analysis and as shown in the figure below, the abnormal signals are captured.
RIGOL 10 Waveform Record Waveform Playback Waveform playback can play back the waveforms currently recorded. Press Record Mode and use to select “Play back”. At this point, the information as shown in the figure below is displayed at the upper right corner of the screen, indicating the specific frame currently displayed on the screen. During the playback, this value would change continuously. Please refer to the following descriptions to set the playback parameters. 1.
10 Waveform Record RIGOL 5. End Frame Press this softkey and use to set the end frame of the playback. The default is the total number of frames of the waveform to be played back. 6. Playback Operation Waveform playback can be realized through the menu or the shortcut buttons at the front panel. Menu Press Operate and select “ start playing back. Front Panel ” to Press Operate and select “ ” to pause the playback.
10 Waveform Record RIGOL Waveform Analysis This function is used to analyze the recorded waveform. Press Record Mode and use to select “Analyze” to open the waveform analysis menu. At this point, the screen is divided into two display areas as shown in the figure below. Settings and Results Templet Display Current Frame Please refer to the following explanations to set the waveform analysis parameters. 1. Analyze Press Analyze to select the desired analysis mode.
10 Waveform Record 3. RIGOL Start Press Satrt to enable waveform analysis. Note that during the analysis process, the progress bar is displayed and the parameters can not be modified. After the analysis finishes, the analysis results of “Error Frames”, “Current Error” and “CurFrame Diff” are displayed; at the same time, the first error frame is located as shown in the figure below.
RIGOL 10 Waveform Record 4. Cancel During the analysis, users can press Cancel to stop the analysis and press Start again to restart the analysis. 5. Previous After the waveform analysis finishes, pressing Previous can locate the error frame previous to the current error frame. Pressing RUN/STOP can also perform the operation. 6. Next After the waveform analysis finishes, pressing Next can locate the error frame following the current error frame. Pressing SINGLE can also perform the operation.
10 Waveform Record RIGOL than the threshold currently set. 9. TempletDisp Press TempletDisp to enable or disable templet display. For the templet setting, refer to the introductions in “Analysis Based on Trace” and “Analysis Based on Pass/Fail Mask”.
RIGOL 10 Waveform Record Analysis Based on Trace Press Analyze and select “Trace”. Then, set the templet used in analysis based on trace through the method below. 1. Trace Press Trace to select the creation method of analysis templet. Current Frame: select the current frame as the analysis templet. Average: select the average of all the frames of this analysis as the analysis templet. 2. Setup Templet Press Setup Templet to configure the templet immediately.
10 Waveform Record RIGOL Analysis Based on Pass/Fail Mask Press Analyze and select “Pass/Fail”. Then, set the templet used in analysis based on Pass/Fail mask through the method below. 1. X Mask Press this softkey and use to adjust the horizontal threshold and the range is from 0.02 div to 4.00 div. During the adjustment, two curves would be displayed to show the outline of the current mask as shown in the figure below. Outline of the Current Mask 2.
11 Display Control RIGOL 11 Display Control You can set the type, persistence time and brightness of waveform display as well as the grid type, grid brightness of the screen display and the menu display time.
11 Display Control RIGOL To Select the Display Type Display Press Display Type to set the waveform display mode to “Vectors” or “Dots”. 11-2 Vectors: the sample points are connected by lines and displayed. Normally, this mode can provide the most vivid waveform to view the steep edge of the waveform (such as square waveform). Dots: display the sample points directly. You can directly view each sample point and use the cursor to measure the X and Y values of the sample point.
11 Display Control RIGOL To Set the Persistence Time Press Display Persis.Time to set the persistence time of the oscilloscope to Min, specific values (From 50 ms to 20 s) or Infinite. In the following part, a frequency sweep signal of the sine waveform is used to demonstrate the waveform effects in different persistence times. 1. Min Enable to view waveform changing in high refresh rate. 2.
RIGOL 3. 11-4 11 Display Control Infinite In this mode, the oscilloscope displays the waveform newly acquired without clearing the waveforms acquired formerly. The waveforms acquired formerly will be displayed in relatively low-brightness color and the waveform newly acquired will be displayed in normal brightness and color. Infinite persistence can be used to measure noise and jitter and to capture incidental events.
11 Display Control RIGOL To Set the Waveform Brightness Press Display WaveIntensity or turn when the menu is hidden to adjust the waveform brightness of the analog channel. The default is 50% and the range available is from 0% to 100%. To Set the Screen Grid Press Display Grid to set the screen grid type. : turn the background grid and coordinate on. : turn the background grid off. : turn the background grid and coordinate off.
12 Store and Recall RIGOL 12 Store and Recall Users can save the current settings, waveforms, and screen image of the oscilloscope in internal memory or external USB mass storage device (such as USB storage device) in various formats and recall the stored settings or waveforms when needed.
12 Store and Recall RIGOL Storage System Press Storage to enter the store and recall setting interface. Storage Insert the USB storage device This oscilloscope provides two USB Host interfaces at the front and rear panels to connect USB storage device for external storage. The USB storage device connected is marked as “Disk D” (front panel) and “Disk E” (rear panel).
12 Store and Recall RIGOL Storage Type Press Storage Storage to select the desired storage type. The default is “Traces”. The storage and recall descriptions of each type are as follows: 1. Traces Save the waveform data in external memory in “*.trc” format. The data of all the channels turned on can be saved in different formats. At recall, the data will be displayed on the screen directly. 2. Waveforms Save the waveform data in external memory in “*.wfm” format.
RIGOL 5. 12-4 12 Store and Recall CSV Save the waveform data displayed on the screen or of the specified channels in external memory in a single “*.csv” file. You can specify the file name and the saving directory and save the corresponding parameter file (*.txt) under the same directory using the same file name. The recall of CVS and parameter files is not supported. After selecting this type: Press DataDepth to select “Displayed” or “Maximum”.
12 Store and Recall RIGOL Internal Storage and Recall Internal storage and recall support “Setups” in Storage. In the following part, the storage and recall method and procedures are introduced. 1. Save the specified oscilloscope setting in internal memory. 1) Connect the signal to the oscilloscope and obtain stable display. 2) Press Storage Storage to select “Setups” and press Save to turn on the interface as shown in Figure a.
RIGOL 12 Store and Recall Figure b Note: in internal storage, New File and New Folder are not available. 2. Load the specified type of file in internal memory. 1) Press Storage Storage to select “Setups” and then press Load to turn on the interface as shown in Figure c. Use to select “Local Disk” and then press down to open the local disk (Figure d). Figure c 2) As shown in Figure d, use to select the desired file to load and press Load to load the file selected.
12 Store and Recall RIGOL External Storage and Recall Before using external storage and recall, make sure that the USB storage device is connected correctly. External storage supports all the types of files in Storage but in recall, “Picture” and “CSV” are not supported. 1. Save the specified type of file in the external USB storage device. 1) Connect the signal to the oscilloscope and obtain stable display.
RIGOL 12 Store and Recall 4) After the storage position is selected, press New File to turn on the interface as shown in Figure g. Refer to the descriptions in “To Create a New File or Folder” to create a new file name. Figure g 5) Press OK to execute the saving operation. 2. Load the specified type of file in the external USB storage device. 1) Press Storage Storage to select “Traces” and then press Load to turn on the interface as shown in Figure h.
12 Store and Recall RIGOL Figure i DS4000 User’s Guide 12-9
RIGOL 12 Store and Recall Disk Management Press Storage Disk.Manage to turn on the disk management interface as shown in the figure below and use to select the desired disk. The disk currently selected is displayed in green and press down to open the disk selected. Execute the following operations through the disk management menu: To Select File Type To Create New, Delete or Rename Files or Folders To Copy and Paste Files.
12 Store and Recall RIGOL To Select File Type Except the file types in Storage, the oscilloscope can also display, save or read some files for advanced applications such as mask file of the Pass/Fail test (*.pf), waveform record file (*.rec), upgrade file (.rgl), parameter file (*.txt) and reference waveform file (*.ref). Press Storage Disk.Manage File Type to select the desired file type. The default is “*.*”.
RIGOL 12 Store and Recall To Create a New File or Folder This operation is only valid in external storage. Before using external disk, make sure that the USB storage device is connected correctly. First, press Storage Disk.Manage and use to select and open the external disk (“Disk D” or “Disk E”). Then, select the desired file type and select the desired directory under which to create a new file or folder. The default is the root directory of the USB storage device.
12 Store and Recall RIGOL English Input Method For example, create a file or folder with the name “Filename”. 1. Press Keyboard. 1) Use to select English input method “En” and uppercase input state “aA”. 2) Use to input the letter “F”. If the input is wrong, press Delete to delete the character input. 3) Use to select lowercase input state “Aa”. 4) Use to input the remaining letters “ilename”. Name Input Area 2. 3.
12 Store and Recall RIGOL Chinese Input Method For example, create a file or folder with the name “文件名”. 1. Press Keyboard. 1) Use to select Chinese input method “中” and lowercase input state “ Aa”. Note that Chinese is added in the menu items at the right of the screen. 2) Use to input the pinyin “wen”. If the input is wrong, press Delete to delete the pinyin input. After “wen” is input, a series of Chinese characters appear in the “Chinese Chraracter Selecting Area”.
12 Store and Recall RIGOL To Delete a File or Folder Folder operation is valid only in external storage. Before using the external disk, make sure that the USB storage device is connected correctly. 1. Delete a file in internal memory. 1) Press Storage Disk.Manage and use to select and open the local disk (“local Disk”). 2) Press File Type to select the desired type of file to delete (note that the file types of internal storage include “*.stp”, “*.ref” and “*.pf”).
RIGOL 12 Store and Recall To Rename a File or Folder Rename operation is valid only in external storage. Before using the external disk, make sure that the USB storage device is connected correctly. Press Storage Disk.Manage and use to select and open the external disk (“Disk D” or “Disk E”). Use to select the desired file or folder to rename and then press Rename to turn on the rename interface. For specific operations, please refer to the descriptions in “To Create a New File or Folder”.
12 Store and Recall RIGOL Factory Press Default to return the oscilloscope to its factory state (refer to the table below).
12 Store and Recall RIGOL Input Impedance 1 MΩ Channel Invert OFF Vertical Scale Coarse Channel Unit V Acquisition Setting (Acquire) Acquisition Mode Normal Sampling Mode Real Time Memory Depth Auto Trigger Setting (TRIGGER) Trigger Type Edge Source CH1 Slope Rising Edge Trigger Mode Auto Trigger Coupling DC Trigger Holdoff 100 ns Display Setting (Display) Display Type Vectors Persistence Time Min Waveform Intensity 50% Screen Grid Brightness 50% Menu Display Infinite C
12 Store and Recall RIGOL Cursor B CH2 CurA -4*1 us CurB 4*1 us Storage Setting (Storage) Storage Type Traces Utility Function Setting (Utility) I/O Setting Network Configuration Mode DHCP, Auto IP USB Device Computer Sound Sound OFF Pass/Fail Test Enable Test OFF Source CH1 Operate OFF X Mask 0.24 div Y Mask 0.
12 Store and Recall RIGOL Aux TrigOut Reference Clock ClockOutput Math Operation Setting (MATH->Operation) Operate OFF A+B Source A CH1 Source B CH1 Invert OFF Vertical Scale 2V A-B Source A CH1 Source B CH1 Invert OFF Vertical Scale 2V A*B Source A CH1 Source B CH1 Vertical Invert OFF Scale 2U A/B Source A CH1 Source B CH1 Invert OFF Vertical Scale 2U FFT Source CH1 Window Function Rectangle Display Split Scale Vrms Vertical Scale 10 dB Vrms/div Horizontal
12 Store and Recall RIGOL Source B CH1 Invert OFF Vertical Scale 1U Advanced Operation Expression OFF Expression CH1+CH2 Vertical Scale 2V Protocol Decoding (Decode 1, Decode 2) Decoding Type Parallel BUS Status OFF Format Hex Offset 0 Threshold 0 Parallel Clock Channel None Slope Rising Edge Bus Bits 1 Current Bit 0 Channel CH1 RS232 TX CH1 RX CH2 Polarity Negative Endian LSB Baud 9600 bps Data Bits 8 Stop Bit 1 Even-Odd Check None Packet OFF Packet End
12 Store and Recall RIGOL MISO Channel CH1 MISO Channel Low MOSI Channel CH1 MOSI Polarity Low MOSI Data Bits 8 MOSI Endian MSB I2C SCLK CH1 SDA CH2 Reference Waveform Setting (REF) Channel Setting REF1 Current Channel REF1 Source CH1 12-22 DS4000 User’s Guide
13 System Function Setting RIGOL 13 System Function Setting The contents of this chapter: Remote Interface Configuration System-related DS4000 User’s Guide 13-1
RIGOL 13 System Function Setting Remote Interface Configuration DS4000 can communicate with PC via LAN and USB buses. Please refer to the introduction below to configure the corresponding interface before using the remote interfaces. LAN Setting Press Utility IO Setting LAN Set to turn on the LAN setting interface. You can view the network connection status and configure the network parameters.
13 System Function Setting RIGOL IP Configuration Type (DHCP) The configuration type of the IP address can be DHCP, auto IP or static IP. In different IP configuration type, the configuration mode of the network parameters (such as the IP address) is different. Press Config Mode and use to select “DHCP”. Then press down to select this type. When DHPC type is valid, the DHCP server will assign the network parameters (such as the IP address) for the oscilloscope.
RIGOL 13 System Function Setting nnn is from 0 to 223 (except 127) and the ranges of the other three nnn are from 0 to 255. You are recommended to ask your network administrator for an IP address available. Press IP Address and use to input the desired IP address. This setting will be saved in the non-volatile memory and if DHCP and Auto IP are set to “Off”, the oscilloscope will load the preset IP address automatically at the next power-on. 2. Set the Subnet Mask. The format of the subnet mask is nnn.
13 System Function Setting RIGOL range of the first nnn is from 0 to 223 (except 127) and the ranges of the other three nnn are from 0 to 255. You are recommended to ask your network administrator for an address available. press DNS and use to input the desired address. Generally, users do not need to set the DNS, therefore this parameter setting can be ignored.
RIGOL 13 System Function Setting USB Device This oscilloscope can communicate with PC or printer via the USB Device interface at the rear panel. You need to set the oscilloscope to make it match with different device types. Press Utility IO Setting USB Device and use device type.
13 System Function Setting RIGOL System-related Sound When the sound is enabled, you can hear the sound of the beeper when you press a function key or a menu softkey or when the prompt message pops up. Press Utility Sound to select sound is turned on, a trumpet icon (on) or (off). The default is off. When the will be displayed at the lower right corner of the screen. Language This oscilloscope supports multiple language menus, Chinese/English help and prompt messages.
RIGOL 13 System Function Setting System Information Press Utility System System Info to view the version information of your oscilloscope. The system information contains the following contents as shown in the figure below. Power-off Recall You can set the system configuration to be recalled when the oscilloscope is powered on again after power-off. Press Utility System Power On to select “Last” (default) or “Default”. Last: return to the setting of the system at last power-off.
13 System Function Setting RIGOL System Time The system time is displayed at the lower right corner of the screen in “hh:mm (hour:minute)” format. When printing or storing a waveform, the file output will contain this time information. Press Utility System System Time System Time to turn on the setting interface as follows: In the figure, the item in green (such as 2011) is the item that can be modified currently. Use to modify and press down to complete the input.
RIGOL 13 System Function Setting Self-test Information Press Utility System SelfTestInfo to view the result of the last self-test of the oscilloscope. The self-test result usually contains the contents as shown in the figure below. Screen When the oscilloscope enters the idle state and holds for a certain period of time, the screen saver program will be enabled. Press Utility System Screensaver to open the screen saver setting menu. “Default” means using the icon RIGOL as the screen saver icon.
13 System Function Setting RIGOL Error Information If this menu is displayed in gray, it means that the circuit board of the oscilloscope is working normally. Otherwise, an exclamation mark will be displayed in the status bar at the lower right corner of the screen, indicating that there is error information currently and at this point, press this key (press Utility System ErrorInfo) to view the error information.
RIGOL 13 System Function Setting Self-calibration The self-calibration program can quickly make the oscilloscope reach the best working state to get the most precise measurement values. You can perform self-calibration at any time especially when the change of the environment temperature is up to or more than 5 ℃. Make sure that the oscilloscope has been warmed up or operated for more than 30 minutes before the self-calibration.
13 System Function Setting RIGOL Power Status Users can set the power status of the oscilloscope after power-on. Connect the oscilloscope to the AC source using the power cord and turn on the power switch at the rear panel. At this point, the oscilloscope is powered on. Press Utility PowerStatus to select “Default” or “Open”. Default: after the oscilloscope is powered on, you need to press the power key at the front panel to start up the oscilloscope.
RIGOL 13 System Function Setting Aux Output Users can set the type of the signal output from the [Trig Out/Calibration] connector at the rear panel. Press Utility AuxOutput to select the desired output type. 1. TrigOut After this type is selected, the oscilloscope outputs a signal that can reflect the current capture rate of the oscilloscope at each trigger. 2.
13 System Function Setting RIGOL Reference Clock This oscilloscope can output the internal 10 MHz sample clock from the [10MHz In/Out] connector at the rear panel and accept an external 10 MHz clock to synchronize multiple oscilloscopes. 10MHz In/Out Press Utility RefClock to select the desired clock type. ClockOutput: configure the [10MHz In/Out] connector as an output to output the internal 10 MHz clock.
13 System Function Setting RIGOL Option Management This oscilloscope provides multiple options to fulfill your measurement requirements. Please contact your RIGOL sales representative or RIGOL technical support to order the corresponding options. You can view the options currently installed on the oscilloscope or activate the newly bought option serial number through this menu. Press Utility Options Installed to view the options currently installed on the oscilloscope.
14 Remote Control RIGOL 14 Remote Control The oscilloscope can be controlled remotely mainly through the following two methods. User-defined programming Users can program and control the oscilloscope by using the SCPI (Standard Commands for Programmable Instruments) commands. For more information about the commands and programming, refer to the Programming Guide.
14 Remote Control RIGOL Remote Control via USB 1. Connect the device Connect the oscilloscope (USB DEVICE) with your PC using a USB cable. 2. Install the USB driver This oscilloscope is a USBTMC device. Assuming that your PC has already been installed with Ultra Sigma, after you connect the oscilloscope to the PC and turn both on for the first time (the oscilloscope is automatically configured to USB interface), the New Hardware Wizard as shown in the figure below is displayed on the PC.
14 Remote Control RIGOL 3 4 5 6 DS4000 User’s Guide 14-3
14 Remote Control RIGOL 7 3. Search device resource Start up the Ultra Sigma and the software will automatically search for the oscilloscope resources currently connected to the PC. You can also click to search the resources. During the search, the status bar of the software is as shown in the figure below: 4.
14 Remote Control RIGOL and the model number and USB interface information of the instrument will also be displayed as shown in the figure below. 5. Communication test Right click the resource name “DS4054 (USB0::0x1AB1::0x04B1::DS4A0000000001::INSTR)” to select “SCPI Panel Control” to turn on the remote command control panel (as shown in the figure below) through which you can send commands and read data.
14 Remote Control RIGOL Remote Control via LAN 1. Connect the device Connect the oscilloscope to your LAN using a network cable. 2. Configure network parameters Configure the network parameters of the oscilloscope according to the description in “LAN Setting”. 3. Search device resource Start up the Ultra Sigma and click figure below is displayed. Click .
14 Remote Control RIGOL Figure (b) Figure (c) 4.
RIGOL 14 Remote Control 5. Communication test Right click the resource name “DS4054 (TCPIP::172.16.3.118::INSTR)” to select “SCPI Panel Control” to turn on the remote command control panel (as shown in the figure below) through which you can send commands and read data. 6. Load LXI webpage As this oscilloscope conforms to LXI-C standards, you can load LXI webpage through Ultra Sigma (right-click the resource name and select LXI-Web; or directly input the IP address in the browser).
15 Troubleshooting RIGOL 15 Troubleshooting The commonly encountered failures and their solutions are listed below. When you encounter those problems, please solve them following the corresponding steps. If the problem remains still, please contact RIGOL and provide your device information (Utility System System Info). 1. The screen is still dark (no display) after power on: (1) Check if the power is correctly connected. (2) Check if the power switch is really on. (3) Check if the fuse is burned out.
RIGOL 15 Troubleshooting signal should use “Video” trigger. Only when the proper trigger type is used, can the waveform be displayed stably. (3) Try to change the Coupling to “HF Reject”or “LF Reject” to filter out the high-frequency or low-frequency noise that disturbs the trigger. (4) Change the trigger holdoff setting. 5. No display after pressing RUN/STOP: Check if the MODE at the trigger panel (TRIGGER) is on “Normal” or “Single” and if the trigger level exceeds the waveform range.
16 Specifications RIGOL 16 Specifications All the specifications are guaranteed except the parameters marked with “Typical” and the oscilloscope needs to operate for more than 30 minutes under the specified operation temperature. Sample Sample Mode Real-time Sample Real Time Sample Rate 4.0 GSa/s (single-channel) 2.
16 Specifications RIGOL Maximum Input Voltage (1MΩ) Maximum Input Voltage of the Analog Channel CAT I 300 Vrms, CAT II 100 Vrms, Transient Overvoltage 1000 Vpk with RP2200 10:1 probe: CAT II 300 Vrms with RP3300 10:1 probe: CAT II 300 Vrms with RP3500 10:1 probe: CAT II 300 Vrms with RP5600 10:1 probe: CAT II 300 Vrms Horizontal Timebase Scale DS405x/ DS403x: 1 ns/div to 50 s/div DS402x: 2 ns/div to 50 s/div DS401x: 5 ns/div to 50 s/div Timebase Accuracy ≤ ±4ppm Clock Drift ≤ ±2 ppm/year Delay Ran
16 Specifications RIGOL 1 mV/div to 1 V/div (50 Ω) Offset Range 1 mV/div to 124 126 mV/div to 1 1 mV/div to 225 230 mV/div to 5 Dynamic Range mV/div: ± 1.2V (50 Ω) V/div: ± 12V (50 Ω) mV/div: ± 2V (1MΩ) V/div: ± 40V (1MΩ) ± 5 div 2 Bandwidth Limit Low Frequency Response (AC Coupling -3dB) DS405x/ DS403x: 20 MHz/100 MHz/200 MHz DS402x: 20 MHz/100 MHz DS401x: 20 MHz ≤5 Hz (on BNC) Calculated Rise Time2 DS405x: DS403x: DS402x: DS401x: 700 ps 1 ns 1.8 ns 3.
16 Specifications RIGOL Pulse Trigger Pulse Condition Positive Pulse Width (greater than, lower than, within specific interval) Negative Pulse Width (greater than, lower than, within specific interval) Pulse Width Range 4 ns to 4 s Slope Trigger Slope Condition Positive Slope (greater than, lower than, within specific interval) Negative Slope (greater than, lower than, within specific interval) Time Setting 10 ns to 1 s Video Trigger Signal Standard Support standard NTSC, PAL and SECAM broadcasti
16 Specifications RIGOL Trigger Condition SOF, EOF, Frame Type, Frame Error Baud Rate 10kbps, 20kbps, 33.3kbps, 50kbps, 62.5kbps, 83.
16 Specifications RIGOL Math Operation Waveform Operation A+B, A-B, A× B, A/B, FFT, Editable Advanced Operation, Logic Operation FFT Window Function Rectangle, Hanning, Blackman, Hamming FFT Display Split, Full Screen FFT Vertical Scale Linear RMS, dBV RMS Logic Operation AND, OR, NOT, XOR Math Function Intg, Diff, Log, Exp, Sqrt, Sine, Cosine, Tangent Number of Buses for Decoding 2 Decoding Type Parallel (standard), RS232/UART (option), I2C (option) , SPI (DS4XX4 option), CAN (option), Fle
16 Specifications RIGOL General Specifications Probe Compensation Output Output Voltage2 Frequency 2 About 3 V, peak-peak 1 kHz Power Power Voltage 100 to 127 V, 45 to 440Hz 100 to 240 V, 45 to 65Hz Power Maximum 120W Fuse 3 A, T Degree, 250 V Environment Temperature Range Operating: 0 ℃ to +50 ℃ Non-operating: -20 ℃ to +70 ℃ Cooling Method Humidity Range fan cooling Under +35 ℃: ≤90% Relative Humidity +35 ℃ to +50 ℃: ≤60% Relative Humidity Altitude Operating: under 3,000 meters Non-operatin
17 Appendix RIGOL 17 Appendix Appendix A: Accessories and Options Model Standard Accessories Optional Accessories Decoding Options Description Order Number DS4054 (500 MHz, 4-channel) DS4054 DS4052 (500 MHz, dual-channel) DS4052 DS4034 (350 MHz, 4-channel) DS4034 DS4032 (350 MHz, dual-channel) DS4032 DS4024 (200 MHz, 4-channel) DS4024 DS4022 (200 MHz, dual-channel) DS4022 DS4014 (100 MHz, 4-channel) DS4014 DS4012 (100 MHz, dual-channel) DS4012 Power Cord conforming to the standard of
RIGOL 17 Appendix Appendix B: Warranty RIGOL warrants that its products mainframe and accessories will be free from defects in materials and workmanship within the warranty period. If a product is proven to be defective within the respective period, RIGOL guarantees the free replacement or repair of products which are approved defective. To get repair service, please contact with your nearest RIGOL sales and service office.
17 Appendix RIGOL Appendix C: Any Comment or Question? If you have any question or comment on our document, please mail to: service@rigol.
Index RIGOL Index - Duty ...................................... 6-16 - Width .................................... 6-16 + Duty..................................... 6-16 + Width ................................... 6-16 Acquisition Mode ........................ 4-2 Addition ..................................... 6-2 AND .......................................... 6-8 Antialiasing ................................ 4-9 Auto .......................................... 5-4 Auto IP ...................................
RIGOL Pulse Width Setting................... 5-11 Record Constant On .......... 10-1, 10-4 Rectangle .................................. 6-6 Reference Clock........................ 1-13 Rise Time ................................ 6-16 Roll ........................................... 3-8 RS232 Decoding ......................... 7-5 Sample Rate .............................. 4-6 Setups ..................................... 12-3 Single ........................................ 5-4 Slope Condition ...............