Datasheet

RIGOL
4
Specications
All the specications are guaranteed except the parameters marked with "Typical" and the oscilloscope needs to operate for more than 30
minutes under the specied operation temperature.
Sample
Sample Mode Real-time Sampling
Real-time Sample
Rate
1 GSa/s on both channels
Peak Detection 500 ps
Averaging
After all the channels have reached N times of sampling at the same time, N can be 2, 4, 8, 16, 32, 64, 128, 256,
512, 1024, 2048, 4096, or 8192.
High Resolution 12-bit resolution when ≥5 μs/div @ 1 GSa/s
Memory Depth Auto, 7 kpts, 70 kpts, 700 kpts, 7 Mpts, and 28 Mpts
Input
Number of Channels 2 analog channels
Input Coupling DC, AC or GND
Input Impedance (1 MΩ±1%)||(16 pF±3 pF) or 50 Ω±1.5%
Probe Attenuation
Coefcient
0.01X-1000X, at 1-2-5 step
Maximum Input
Voltage (1 MΩ)
CAT I 300 Vrms, CAT II 100 Vrms, Transient Overvoltage 1000 Vpk
Horizontal
Timebase Scale
DS2102E: 5.000 ns/div to 1.000 ks/div
DS2202E: 2.000 ns/div to 1.000 ks/div
Channel to Channel
Skew
1 ns (typical), 2 ns (maximum)
Max. Record Length 28 Mpts on both channels
Timebase Accuracy
[1]
≤±25 ppm
Clock Drift ≤±5 ppm/year
Max. Delay Range
Negative Delay: ≥1 screen width
Positive Delay: 1 s to 100 ks
Timebase Mode Y-T, X-Y, Roll
Number of X-Ys 1 path
Waveform Capture
Rate
[2]
50,000 wfms/s (dots display)