User manual
5 To Trigger the Oscilloscope RIGOL
DS2000A User’s Guide
5-11
Pulse Width Setting:
In this oscilloscope, positive pulse width is defined as the time difference between
the two crossing points of the trigger level and positive pulse as shown in the figure
below.
When the Pulse Condition is set to , , or , press
Setting and use to input the desired value. The range available is from 2 ns
to 4 s.
When the Pulse Condition is set to or , press Upper Limit
and Lower Limit and use to input the desired values respectively. The
range of the upper limit is from 10 ns to 4 s. The range of the lower limit is from
2 ns to 3.99 s. Note that the lower limit of the pulse width must be lower than
the upper limit.
Trigger Mode:
Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto,
normal or single. The corresponding status light of the current trigger mode turns on.
Trigger Setting:
Press Setting to set the trigger parameters (trigger holdoff and noise rejection)
under this trigger type.
Trigger Level:
Use TRIGGER LEVEL to modify the level. For details, please refer to the
description of
Trigger Level on page 5-9.
Trigger level
Positive
pulse width
A
B