User manual
RIGOL 17 Specifications
DS2000A User’s Guide
17-10
General Specifications
Probe Compensation Output
Output Voltage
1
About 3 V, peak-peak
Frequency
1
1 kHz
Power
Power Voltage 100 to 240 V, 45 to 440 Hz
Power Maximum 50 W
Fuse 2 A, T Degree, 250 V
Environment
Temperature Range O p e ra tin g : 0 ℃ to + 5 0 ℃
Non-operating: -40 ℃ to + 7 0 ℃
Cooling Method Fan cooling
Humidity Range
0℃ to +30℃: ≤95% Relative Humidity
+30℃ to +40℃: ≤75% Relative Humidity
+40℃ to +50℃: ≤45% Relative Humidity
Altitude Operating: under 3,000 meters
Non-operating: under 15,000 meters
Physical Characteristics
Size
3
Width×Height×Depth = 361.6 mm× 179.6 mm×130.8
mm
Weight
4
Package Excluded 3.9 kg ± 0.2 kg
Package Included 4.5 kg ± 0.5 kg
Calibration Interval
The recommended calibration interval is one year.
Regulatory Information
Electromagnetic
Compatibility
2004/108/EC
Execution standard EN 61326-1:2006 EN 61326-2-1:2006
Safety UL 61010-1:2004; CAN/CSA-C22.2 NO. 61010-1-2004;
EN 61010-1:2001; IEC 61010-1:2001
1 Typical.
2 Maximum value. 20ns, single-channel mode, dots display, auto memory depth.
3 Supporting legs and handle folded, knob height included.
4 Standard configuration.