Specifications
2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-53
:CALCulate:MULTiply:VSCale
Syntax
:CALCulate:MULTiply:VSCale <scale>
:CALCulate:MULTiply:VSCale?
Description
Set or query the vertical scale of the multiplication operation result.
Parameter
Name Type Range Default
<scale> Real
The default range is from 50nU to 10MU.
The actual range is related to the vertical scale of the signal
source channel.
2U
Note: For the channel vertical scale, refer to the :CHANnel<n>:SCALe command.
Return Format
The query returns the vertical scale of the multiplication operation result in scientific notation.
:CALCulate:MULTiply:VOFFset
Syntax
:CALCulate:MULTiply:VOFFset <offs>
:CALCulate:MULTiply:VOFFset?
Description
Set or query the vertical offset of the multiplication operation result.
Parameter
Name Type Range Default
<offs> Real -40×VScale to 40×VScale 0
Note: For the VScale (the vertical scale of MATH), refer to the :CALCulate:MULTiply:VSCale command.
Return Format
The query returns the vertical offset of the multiplication operation result in scientific notation.