Specifications

2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-51
:CALCulate:SUB:VSCale
Syntax
:CALCulate:SUB:VSCale <scale>
:CALCulate:SUB:VSCale?
Description
Set or query the vertical scale of the subtraction operation result. The unit is related to the amplitude unit
of the signal source channel.
Parameter
Name Type Range Default
<scale> Real
The default range is from 1p to 200.
The actual range is related to the vertical scale of the
signal source channel.
2V
Note: For the channel vertical scale, refer to the :CHANnel<n>:SCALe command.
Return Format
The query returns the vertical scale of the subtraction operation result in scientific notation.
:CALCulate:SUB:VOFFset
Syntax
:CALCulate:SUB:VOFFset <offs>
:CALCulate:SUB:VOFFset?
Description
Set or query the vertical offset of the subtraction operation result. The unit is related to the amplitude unit
of the signal source channel.
Parameter
Name Type Range Default
<offs> Real -40×VScale to 40×VScale 0
Note: For the VScale (the vertical scale of MATH), refer to the :CALCulate:SUB:VSCale command.
Return Format
The query returns the vertical offset of the subtraction operation result in scientific notation.