Specifications

2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-43
:BUS<n>:CAN:STYPe
Syntax
:BUS<n>:CAN:STYPe <stype>
:BUS<n>:CAN:STYPe?
Description
Set or query the signal type of CAN decoding on the specified bus.
Parameter
Name Type Range Default
<n> Discrete {1|2} --
<stype> Discrete {TX|RX|CANH|CANL|DIFFerential} RX
Return Format
The query returns TX, RX, CANH, CANL or DIFF.
:BUS<n>:CAN:BAUD
Syntax
:BUS<n>:CAN:BAUD <baud>
:BUS<n>:CAN:BAUD?
Description
Set or query the signal rate of CAN decoding on the specified bus. The default unit is b/s.
Parameter
Name Type Range Default
<n> Discrete {1|2} --
<baud> Discrete {100000|125000|250000|400000|500000|800000|1000000|USER} 500000
Note: When the signal rate is set to USER, you can use the :BUS<n>:CAN:BUSer command to set a user-defined
signal rate.
Return Format
The query returns the signal rate of CAN decoding on the specified bus in integer or returns USER.