Specifications
2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-351
:WAVeform:YINCrement?
Syntax
:WAVeform:YINCrement?
Description
Query the voltage value per unit of the channel source currently selected (refer to
the
:WAVeform:SOURce command) in the Y direction. The unit is the same with the unit of the signal
source.
Explanation
YINCrement=VerticalScale/25
Return Format
The query returns the voltage value per unit of the channel source currently selected in the Y direction in
scientific notation.
:WAVeform:YORigin?
Syntax
:WAVeform:YORigin?
Description
Query the vertical offset relative to the vertical reference position (refer to the
:WAVeform:YREFerence?
command) of the channel source currently selected (refer to the
:WAVeform:SOURce command) in the Y
direction. The unit is the same with the unit of the signal source.
Return Format
The query returns the vertical offset relative to the vertical reference position of the channel source
currently selected in the Y direction in integer.
:WAVeform:YREFerence?
Syntax
:WAVeform:YREFerence?
Description
Query the vertical reference position of the channel source currently selected (refer to
the
:WAVeform:SOURce command) in the Y direction.
Explanation
YREFerence is always 127 (the screen bottom is 0 and the screen top is 255).
Return Format
The query returns 127.