Specifications
RIGOL 2 Command System
2-196 MSO2000A/DS2000A Programming Guide
:MEASure:VMAX:SMAXimum
:MEASure:VMAX:SMINimum
:MEASure:VMAX:SCURrent
:MEASure:VMAX:SAVerage
:MEASure:VMAX:SDEViation
Syntax 1
:MEASure:VMAX:SMAXimum <chan>
:MEASure:VMAX:SMINimum <chan>
:MEASure:VMAX:SCURrent <chan>
:MEASure:VMAX:SAVerage <chan>
:MEASure:VMAX:SDEViation <chan>
Description 1
Enable the amplitude maximum value measurement function of the specified channel and the statistic
function of the measurement results.
Syntax 2
:MEASure:VMAX:SMAXimum? <chan>
:MEASure:VMAX:SMINimum? <chan>
:MEASure:VMAX:SCURrent? <chan>
:MEASure:VMAX:SAVerage? <chan>
:MEASure:VMAX:SDEViation? <chan>
Description 2
Query the statistic maximum value, minimum value, current value, average value or standard deviation of
the amplitude maximum value measurement results when the amplitude maximum value measurement
function of the specified channel and the statistic function of the measurement results are enabled.
Parameter
Name Type Range Default
<chan> Discrete {CHANnel1|CHANnel2|MATH} --
Explanation
The unit is determined by the unit selected for the signal source channel (refer to
the
:CHANnel<n>:UNITs command).
Return Format
The query returns the statistic maximum value, minimum value, current value, average value or standard
deviation of the amplitude maximum value measurement results in scientific notation. When the
measurement results are invalid, the query returns 9.9e37.