Specifications
RIGOL 2 Command System
2-178 MSO2000A/DS2000A Programming Guide
:MEASure:PWIDth:SMAXimum
:MEASure:PWIDth:SMINimum
:MEASure:PWIDth:SCURrent
:MEASure:PWIDth:SAVerage
:MEASure:PWIDth:SDEViation
Syntax 1
:MEASure:PWIDth:SMAXimum <chan>
:MEASure:PWIDth:SMINimum <chan>
:MEASure:PWIDth:SCURrent <chan>
:MEASure:PWIDth:SAVerage <chan>
:MEASure:PWIDth:SDEViation <chan>
Description 1
Enable the positive pulse width measurement function of the specified channel and the statistic function
of the measurement results.
Syntax 2
:MEASure:PWIDth:SMAXimum? <chan>
:MEASure:PWIDth:SMINimum? <chan>
:MEASure:PWIDth:SCURrent? <chan>
:MEASure:PWIDth:SAVerage? <chan>
:MEASure:PWIDth:SDEViation? <chan>
Description 2
Query the statistic maximum value, minimum value, current value, average value or standard deviation of
the positive pulse width measurement results when the positive pulse width measurement function of the
specified channel and the statistic function of the measurement results are enabled.
Parameter
Name Type Range Default
<chan> Discrete
{D0|D1|D2|D3|D4|D5|D6|D7|D8|D9|D10|D11|
D12|D13|D14|D15|CHANnel1|CHANnel2|MATH}
--
Return Format
The query returns the statistic maximum value, minimum value, current value, average value or standard
deviation of the positive pulse width measurement results in scientific notation. When the measurement
results are invalid, the query returns 9.9e37.