User`s guide

RIGOL 4 To Set the Sample System
DS2000 User’s Guide
4-8
Memory Depth
Memory depth refers to the number of waveform points that the oscilloscope can
store in a single trigger sample and it reflects the storage ability of the sample
memory. DS2000 provides up to 56 Mpts memory depth (option) and 14 Mpts
standard memory depth.
T
Trigger Point
Pre-sample Delayed Sample
Memory Depth
The relation of memory depth, sample rate and waveform length fulfills the equation
below:
Memory depth = sample rate (Sa/s) × waveform length (s/div × div)
Press Acquire Mem Depth, use to switch to the desired memory depth (the
default is auto) and then press down the knob to select the option. You can also
press Mem Depth continuously to switch the memory depth.
When a single channel is enabled, the memory depths available include Auto,
14kPoints, 140kPoints, 1.4MPoints, 14MPoints and 56MPoints (option). In Auto
mode, the oscilloscope selects the memory depth automatically according to the
current sample rate.
When CH1 and CH2 are enabled, the memory depths available include Auto, 7kPoints,
70kPoints, 700kPoints, 7MPoints and 28MPoints (option). In “Auto” mode, the
oscilloscope selects the memory depth automatically according to the current sample
rate.