User`s guide

3 To Set the Horizontal System RIGOL
DS2000 User’s Guide
3-5
X-Y Mode
In this mode, the oscilloscope changes the two channels from voltage-time display
mode to voltage-voltage display mode. Wherein, the X axis and Y axis tracks the
voltages of CH2 and CH1 respectively. The phase deviation between two signals with
the same frequency can be easily measured via Lissajous method. The figure below
shows the measurement schematic diagram of the phase deviation.
According to sinθ=A/B or C/D (wherein, θ is the phase deviation angle between
the two channels and the definitions of A, B, C and D are as shown in the figure
above), the phase deviation angle is obtained, that is:
θ=±arcsin (A/B) or ±arcsin (C/D)
If the principal axis of the ellipse is within quadrant I and III, the phase deviation
angle obtained should be within quadrant I and IV, namely within (0 to π/2) or (3π/2
to 2π). If the principal axis of the ellipse is within quadrant II and IV, the phase
deviation angle obtained should be within quadrant II and III, namely within (π/2 to
π) or (π to 3π/2).
X-Y function can be used to measure the phase deviation occurred when the signal
under test passes through a circuit network. Connect the oscilloscope to the circuit to
II I
III IV
A
B
C
D
The signal must be
centered horizontally
II I
III IV