User`s guide

RIGOL 1 Quick Start
DS2000 User’s Guide
1-10
Probe Compensation
When the probes are used for the first time, you should compensate the probes to
match the input channels of the oscilloscope. Non-compensated or poorly
compensated probes may cause measurement inaccuracy or error. The probe
compensation procedures are as follows.
1. Perform steps 1, 2, 3 and 4 of Function Inspectionin the previous section.
2. Check the waveforms displayed and compare them with the following.
Figure 1-8 Probe Compensation
3. Use a nonmetallic driver to adjust the low-frequency compensation adjustment
hole on the probe until the waveform displayed is as the Perfectly
compensatedin the figure above.
Over compensated Perfectly compensated Under compensated