User`s guide
16 Specifications RIGOL
DS2000 User’s Guide
16-1
16 Specifications
All the specifications are guaranteed except the parameters marked with “Typical”
and the oscilloscope needs to operate for more than 30 minutes under the specified
operation temperature.
Sample
Sample Mode Real-time Sample
Real Time Sample
Rate
2 GSa/s (single-channel)
1 Gsa/s (dual-channel)
Peak Detect
500 ps (single-channel)
1 ns (dual-channel)
Averaging
After both the channels finish N samples at the same time, N
can be 2, 4, 8, 16, 32, 64, 128, 256, 512, 1024, 2048, 4096
or 8192.
High Resolution
12 bits of resolution when ≥5 μs/div @ 1 GSa/s (or ≥10
μs/div @ 500 MSa/s).
Memory Depth
single-channel: Auto, 14k pts, 140k pts, 1.4M pts, 14M pts
and 56M pts (option) are available
dual-channel: Auto, 7k pts, 70k pts, 700k pts, 7M pts and
28M pts (option) are available
Input
Number of Channels Two channels
Input Coupling DC, AC or GND
Input Impedance (1 MΩ±1%) || (16 pF±3 pF)
Probe Attenuation
Coefficient
0.01X to 1000X, in 1-2-5 step
Maximum Input
Voltage (1MΩ)
Maximum Input Voltage of the Analog Channel
CAT I 300 Vrms, CAT II 100 Vrms,