User`s guide

RIGOL 10 Waveform Record
DS2000 User’s Guide
10-12
Analysis Based on Trace
Press Analyze and select “Trace”. Then, set the templet used in analysis based on
trace through the method below.
1. Trace
Press Trace to select the creation method of analysis templet.
Current Frame: select the current frame as the analysis templet.
Average: select the average of the current data frame as the analysis
templet.
2. Setup Template
Press Setup Template to configure the template immediately. After the
analysis starts, the oscilloscope compares each frame with the templet and
determines whether error frame exists according to the threshold currently set.
3. Template Display
Press TemplateDisp to enable or disable template display.