13) Press MENU Vrms at the left of the screen to enable the root mean square
measurement function of the oscilloscope. Read and record Vrms3.
14) Calculate the amplitude loss and compare it to the specification: Amplitude Loss (dB) =
20×lg
(Vrms3/Vrms1)
. Amplitude loss should be lower than -3 dB at this point.
15) Keep other settings of the oscilloscope unchanged and set the vertical scale to 200 mV/div.
16) Output a sine waveform with 1 MHz frequency and 1.2 Vpp amplitude from Fluke 9500B.
17) Repeat step 6).
18) Output a sine waveform with 100 MHz frequency and 1.2 Vpp amplitude from Fluke 9500B.
19) Repeat step 8), 9) and 10).
20) Output a sine waveform with 200 MHz frequency and 1.2 Vpp amplitude from Fluke 9500B.
21) Repeat step 12), 13) and 14).
22) Keep other settings of the oscilloscope unchanged and set the vertical scale to 500 mV/div.
23) Output a sine waveform with 1 MHz frequency and 3 Vpp amplitude from Fluke 9500B.
24) Repeat step 6).
25) Output a sine waveform with 100 MHz frequency and 3 Vpp amplitude from Fluke 9500B.
26) Repeat step 8), 9) and 10).
27) Output a sine waveform with 200 MHz frequency and 3 Vpp amplitude from Fluke 9500B.
28) Repeat step 12), 13) and 14).
29) Turn CH1 off. Test CH2 according to the above test steps and record the test results.