Specifications
RIGOL Chapter 2 Performance Verification Test
DS2000 Performance Verification Guide
2-6
1.2 Vpp/3 Vpp amplitude from Fluke 9500B respectively.
14) Repeat steps 7), 8) and 9).
15) Turn CH1 off. Test CH2 according to the above test steps and record the test results.
Test Record Form:
Channel
Vertical Scale
Test Result
Limit
Pass/Fail
CH1
100 mV/div
Vrms1
≥ -3 dB and ≤ 1 dB
Vrms2
Amplitude Loss
[1]
200 mV/div
Vrms1
Vrms2
Amplitude Loss
500 mV/div
Vrms1
Vrms2
Amplitude Loss
CH2
100 mV/div
Vrms1
Vrms2
Amplitude Loss
200 mV/div
Vrms1
Vrms2
Amplitude Loss
500 mV/div
Vrms1
Vrms2
Amplitude Loss
Note
[1]
: amplitude loss (dB) = 20×lg
(Vrms2/Vrms1)
.