Specifications
RIGOL Chapter 2 Performance Verification Test
DS2000 Performance Verification Guide
2-4
d) Calculate the relative error of each scale.
10) Turn CH1 off. Repeat the above test steps to test CH2 and record the test results.
Test Record Form:
Channel
Vertical Scale
Test Result
Limit
Pass/Fail
Vavg1
Vavg2
Calculation Result
[1]
CH1
500 μV/div
≤ 2%
1 mV/div
2 mV/div
5 mV/div
10 mV/div
20 mV/div
50 mV/div
100 mV/div
200 mV/div
500 mV/div
1 V/div
2 V/div
5 V/div
10 V/div
CH2
500 μV/div
1 mV/div
2 mV/div
5 mV/div
10 mV/div
20 mV/div
50 mV/div
100 mV/div
200 mV/div
500 mV/div
1 V/div
2 V/div
5 V/div
10 V/div
Note
[1]
: the calculation formula is |(Vavg1–Vavg2)–(V
out1
-V
out2
)|/Full Scale×100%; wherein, V
out1
and V
out2
are 3 and
-3 times of the current vertical scale respectively.