Specifications

Chapter 2 Performance Verification Test RIGOL
DS2000 Performance Verification Guide
2-3
DC Gain Accuracy Test
Specification:
DC Gain Accuracy ≤2%×Full Scale
Explanation:
Full Scale = 8 div×vertical scale. Relative error of each scale: |(Vavg1Vavg2)–(V
out1
-V
out2
)|/Full Scale×100%≤2%;
otherwise, the test fails. For example, when the vertical scale is 1 V/div, input DC signals with +3 V
DC
and -3 V
DC
voltages respectively, the values of Vavg1 and Vavg2 are +3.06 V and -3.04 V respectively, the relative error is
|(+3.06 V-(-3.04 V))-(+3 V-(-3 V)) |/(1 V/div*8 div) × 100% = 1.25% and the test passes.
Test Device: Fluke 9500B
Test Procedures:
1) Connect the active head of Fluke 9500B to CH1 of the oscilloscope, as shown in figure
below.
2) Enable the Fluke 9500B and set the output impedance to 1 MΩ.
3) Output a DC signal with +1.5 mV
DC
voltage (V
out1
) from Fluke 9500B.
4) Configure the oscilloscope:
a) Press CH1 in the vertical control area (VERTICAL) at the front panel to enable CH1.
b) Rotate VERTICAL SCALE
to set the vertical scale to 500 μV/div.
c) Rotate HORIZONTAL SCALE to set the horizontal time base to 10 μs.
d) Rotate VERTICAL POSITION
to set the vertical position to 0.
e) Press Acquire Acquisition and use to select “Average”. Then, press Averages
and use to set the number of averages to 32.
5) Press MENU Vavg at the left of the screen to enable the average measurement function
of the oscilloscope. Read and record Vavg1.
6) Adjust Fluke 9500B to output a DC signal with -1.5 mV
DC
output voltage (V
out2
).
7) Press MENU Vavg at the left of the screen to enable the average measurement function
of the oscilloscope. Read and record Vavg2.
8) Calculate the relative error of the vertical scale:
|(Vavg1Vavg2)–(V
out1
-V
out2
)|/Full Scale×100%.
9) Keep other settings of the oscilloscope unchanged:
a) Set the vertical scale to 1 mV/div2 mv/div, 5 mV/div, 10 mV/div, 20 mV/div, 50 mV/div,
100 mV/div, 200 mV/div, 500 mV/div, 1 V/div, 2 V/div, 5 V/div and 10 V/div respectively.
b) Adjust the output voltage of Fluke 9500B to ±3 div respectively.
c) Repeat steps 1), 2), 3), 4), 5) and 6) and record the test results.