Specifications
Chapter 2 Performance Verification Test RIGOL
DS2000 Performance Verification Guide
2-1
Chapter 2 Performance Verification Test
This chapter introduces the performance verification test method and procedures of DS2000 series
digital oscilloscope by taking DS2202 as an example. You can perform the following tests in any order.
In this manual, the test device used is Fluke 9500B. You can also use other devices that fulfill the
specification requirements for the test.
Recommended Device List:
Device
Specification
Recommended Model
Oscilloscope
Calibrator
Output range of DC voltage:
1 MΩ: 1 mV to 200 V
50 MΩ:1 mV to 200 V
The rise time of fast edge signal: ≤ 150 ps
Fluke 9500B
Digital Multimeter
The resistance measurement accuracy is
higher than ±0.1% of reading
RIGOL DM3058/3068
Test Cable BNC (male) to Dual-banana Plug Cable --
Signal Generator Frequency Accuracy: ±1 ppm RIGOL DG5000 series
Test Cable BNC (m)-BNC (m) cable --
Note:
1. Make sure that the oscilloscope passes the self-test and self-calibration before executing
performance verification test.
2. Make sure that the oscilloscope has been warmed up for at least 30 minutes before executing
any of the following tests.
3. Please reset the instrument to the factory setting before or after executing any of the following
tests.