Specifications

2 Command System RIGOL
DS2000 Programming Guide
2-545
:TRIGger:USB:PLEVel
Syntax
:TRIGger:USB:PLEVel <level>
:TRIGger:USB:PLEVel?
Description
Set the trigger level of the D+ data line in USB trigger and the unit is the same with
the current amplitude unit.
Query the current trigger level of the D+ data line in USB trigger.
Parameter
Name Type Range Default
<level> Real
± 5× VerticalScale from the screen center -
OFFSet
0
Note:
For the VerticalScale, refer to the
:CHANnel<n>:SCALe command.
For the OFFSet, refer to the
:CHANNel<n>:OFFSet command.
Return Format
The query returns the trigger level in scientific notation.
Example
:TRIGger:USB:PLEVel 0.16
The query returns 1.600000e-01.