Specifications
RIGOL 2 Command System
2-514 DS2000 Programming Guide
2-514
:TRIGger:RS232:PARity
Syntax
:TRIGger:RS232:PARity <parity>
:TRIGger:RS232:PARity?
Description
Set the even-odd check mode in RS232 trigger when the trigger condition is Error or
Check Error.
Query the current even-odd check mode in RS232 trigger when the trigger condition
is Error or Check Error.
Parameter
Name Type Range Default
<parity> Discrete {EVEN|ODD|NONE} NONE
Note: the even-odd check mode can not be set to NONE when the trigger condition is Check Error.
Explanation
To set the trigger condition, refer to the
:TRIGger:RS232:WHEN command.
Return Format
The query returns EVEN, ODD or NONE.
Example
:TRIGger:RS232:PARity EVEN
The query returns EVEN.