2-501
:TRIGger:DURATion:TUPPer
Syntax
:TRIGger:DURATion:TUPPer <NR3>
:TRIGger:DURATion:TUPPer?
Description
Set the upper limit of the duration in duration trigger and the unit is s.
Query the current upper limit of the duration in duration trigger.
Parameter
Name Type Range Default
<NR3> Real 2ns to 4s 2μs
Note: when the trigger condition is GLESs, the range is from 12ns to 4s.
Explanation
This command is available when the trigger condition (refer to
the
:TRIGger:DURATion:WHEN command) is set to LESS or GLESs.
Return Format
The query returns the upper limit of the duration in scientific notation.
Example
:TRIGger:DURATion:TUPPer 0.000003