Specifications
2 Command System RIGOL
DS2000 Programming Guide
2-453
:TRIGger:RUNT:BLEVel
Syntax
:TRIGger:RUNT:BLEVel <level>
:TRIGger:RUNT:BLEVel?
Description
Set the lower limit of the trigger level in runt trigger and the unit is the same with the
current amplitude unit.
Query the current lower limit of the trigger level in runt trigger.
Parameter
Name Type Range Default
<level> Real
± 5 × VerticalScale from the screen center -
OFFSet
0
Note:
For the VerticalScale, refer to the
:CHANnel<n>:SCALe command.
For the OFFSet, refer to the
:CHANNel<n>:OFFSet command.
Return Format
The query returns the lower limit of the trigger level in scientific notation.
Example
:TRIGger:RUNT:BLEVel 0.16
The query returns 1.600000e-01.