Specifications
2 Command System RIGOL
DS2000 Programming Guide
2-433
:TRIGger:HOLDoff
Syntax
:TRIGger:HOLDoff <value>
:TRIGger:HOLDoff?
Description
Set the trigger holdoff time and the unit is s.
Query the current trigger holdoff time.
Parameter
Name Type Range Default
<value> Real 100ns to 10s 100ns
Explanation
This setting is not available for the Nth edge trigger, video trigger, RS232 trigger, IIC
trigger, SPI trigger and USB trigger.
Return Format
The query returns the trigger holdoff time in scientific notation.
Example
:TRIGger:HOLDoff 0.0000002
The query returns 2.000000e-07.