Specifications
RIGOL 2 Command System
2-410 DS2000 Programming Guide
2-410
:SYSTem:AOUTput
Syntax
:SYSTem:AOUTput <aux output>
:SYSTem:AOUTput?
Description
Set the output signal type of the [Trigger Out] connector at the rear panel.
Query the current output signal type of the [Trigger Out] connector at the rear
panel.
Parameter
Name Type Range Default
<aux output> Discrete {TOUT|PFAil} TOUT
Explanation
TOUT: when this type is selected, the oscilloscope outputs a signal that can reflect
the current capture rate of the oscilloscope at each trigger.
PFAil: when this type is selected, the oscilloscope outputs a pulse signal when failed
waveforms are detected. Connect this signal to other control systems to conveniently
view the test result.
Return Format
The query returns TOUT or PFA.
Example
:SYSTem:AOUTput PFA
The query returns PFA.