2-407
:SYSTem:SSAVer:TIME
Syntax
:SYSTem:SSAVer:TIME <time>
:SYSTem:SSAVer:TIME?
Description
Set the screen saver time.
Query the current screen saver time.
Parameter
Name Type Range Default
<time> Discrete
{1MIN|2MIN|5MIN|15MIN|30MIN|45MIN
|60MIN|2HOur|5HOur|OFF}
OFF
Explanation
When the oscilloscope enters the idle state and holds for a certain period of time, the
screen saver program will be enabled. When OFF is selected, the screen saver
program will not be used.
Return Format
The query returns the current screen saver time or OFF.
Example
:SYSTem:SSAVer:TIME 2MIN