Specifications

RIGOL 2 Command System
2-336 DS2000 Programming Guide
2-336
:MEASure:RTIMe:SCURrent?
Syntax
:MEASure:RTIMe:SCURrent? <chan>
Description
Query the statistic current value of the rise time measurement of the selected
waveform and the unit is s.
Parameter
Name Type Range Default
<chan> Discrete {CHANnel1|CHANnel2} CHANnel2
Return Format
The query returns the measurement result in scientific notation.
Example
:MEASure:RTIMe:SCURrent? CHANnel1
The query returns 3.120000e-04.