Specifications
RIGOL 2 Command System
2-332 DS2000 Programming Guide
2-332
:MEASure:PWIDth:SDEViation?
Syntax
:MEASure:PWIDth:SDEViation? <chan>
Description
Query the statistic standard deviation of the positive pulse width measurement of the
selected waveform and the unit is s.
Parameter
Name Type Range Default
<chan> Discrete {CHANnel1|CHANnel2} CHANnel2
Return Format
The query returns the measurement result in scientific notation.
Example
:MEASure:PWIDth:SDEViation? CHANnel1
The query returns 8.243160e-06.